# How are leaks found on an oxygen cylinder filling ramp?

**How are leaks found and fixed on the high-pressure filling ramp of an on-site oxygen cylinder filling station?**

WHO's Foundations of medical oxygen systems (2023) names leaks from damaged hoses, flexible pigtails, pipes, valves and poor connections as common causes of oxygen fires and says newly assembled equipment should be leak checked; for cylinders on hospital manifolds, HTM 02-01 (2026) finds leaks by listening or with an oxygen-compatible leak-detection fluid.

## Where the leaks are

WHO's technical specifications of 31 December 2024 describe a cylinder filling station as a booster compressor and a filling manifold, also known as a filling ramp, made of a rack, a header, pressure indicators, cylinder connection points and safety chains, with flexible hoses or metallic pigtails intended for high-pressure oxygen filling, a check valve recommended at each cylinder connection, and oxygen-compatible piping from the booster to the header (ISO 15001 or equivalent). WHO's Foundations (2023) says oxygen at high pressure can react violently with oil and grease, and that a leak in a poorly ventilated room can quickly raise the oxygen concentration in the air to a dangerous level.

## Checking for leaks

- After assembly: WHO (2023) says newly assembled equipment should be leak checked and all equipment and hose connections properly fitted; for maintenance, it says to use only tools and substances recommended by the manufacturer and only lubricants and tapes made for oxygen service.
- Matching parts: the Global Fund post-commissioning form (12 September 2025) checks, in its cylinder-filling section, that the cylinder valves are compatible with the pigtails.
- At each cylinder connection: HTM 02-01 does not describe filling ramps. For cylinders connected to a hospital supply manifold, Part B (9.171) says to check for leaks between the cylinder valve and the manifold, usually by listening, and if in doubt with the correct oxygen-compatible leak-detection fluid, wiping off the excess with a clean damp cloth. It also asks for the seal in the cylinder yoke or the O-ring on the bull-nose connection to be inspected for wear, for one seal per connection and for no lubricant or sealing compound.
- Never with a flame: a naked flame or lighted cigarette should not be used to detect leaks, and only oxygen-safe leak-detection fluids should be used (HTM 02-01 Part B, precautions).

## Fixing a leak

- Leaking connection: for supply manifolds, HTM 02-01 Part B (9.171) says to close all cylinders on that bank, disconnect the leaking tailpipe, recheck the seals, replace them if necessary and reconnect.
- Leaking cylinder valve: gas leaking from the valve spindle means the cylinder should be removed from service and returned as faulty, and gland nuts should not be tightened, as this may damage the valve (HTM 02-01 Part B, 9.166 and 9.167). WHO's 2023 poster says not to attempt to repair a cylinder or valve if leakage is detected and suggests replacing damaged components.
- Reopening: cylinder valves should be opened slowly, because ignition may occur by adiabatic compression, particularly with oil, grease or hand-cream residues (HTM 02-01 Part B, 9.164).

Leak testing of the hospital pipeline itself, with pressure decay tests and ultrasonic detectors, is covered in [how leaks are found in a medical gas network](https://library.ultracontrolo.com/answers/how-to-detect-leaks-in-a-medical-gas-network/), and booster maintenance in what maintenance an oxygen booster needs.

### Can ordinary soapy water be used on oxygen fittings?

HTM 02-01 Part B (2026) says only oxygen-safe leak-detection fluids should be used, and excess spray should be wiped off with a clean damp cloth to avoid contaminating the fittings.

### Should a leaking cylinder valve be tightened?

No. HTM 02-01 Part B (2026) says a cylinder leaking from the valve spindle should be removed from service and returned as faulty, and gland nuts should not be tightened.

## Sources

- [WHO, Technical specifications for health facility based medical oxygen systems (31 December 2024), section 1.2 Cylinder filling station, lines 16 to 18](https://iris.who.int/server/api/core/bitstreams/ff437601-add3-44c8-8149-b85f0cdb2b52/content)
- [WHO, Foundations of medical oxygen systems (17 February 2023), section 3.6 and poster B](https://iris.who.int/server/api/core/bitstreams/4ad6f5fd-089e-4142-8439-b64e98d816d8/content)
- [NHS England, HTM 02-01 Medical gas pipeline systems, Part B (2026 edition, August 2026), paragraphs 9.164 to 9.171 and precautions](https://www.england.nhs.uk/wp-content/uploads/2016/05/health-technical-memorandum-02-01-medical-gas-pipeline-systems-part-b-operational-management-aug-26.pdf)
- [Global Fund / Build Health International, Post-Commissioning Assessment Form for Medical Oxygen Plants (12 September 2025), cylinder-filling system](https://resources.theglobalfund.org/media/pt5cp0u5/cr_c19rm-post-commissioning-assessment_form_en.pdf)

Reviewed: 2026-11-18. Ultra Controlo engineering. https://library.ultracontrolo.com/answers/how-are-leaks-found-on-a-cylinder-filling-ramp/