# What does Kenya's 2024 oxygen specification ask of a PSA plant?

**What does Kenya's national 2024 technical specification for medical oxygen ask of a PSA plant?**

Kenya's Ministry of Health 2024 technical specification for medical oxygen devices describes a PSA plant that makes medical oxygen of 93%±3 from ambient air, at a continuous output pressure of 400-600 kPa, with a feed air compressor of at least 750 kPa.

This answer summarises the PSA plant section of the Ministry of Health document, read on 19 November 2026 from a copy hosted at stoppneumonia.org. The PDF carries the Ministry of Health cover and the date 2024; its PSA sections are marked version 1, November 2023. I did not find the document on the Ministry's own website, so a buyer should check the current version with the Ministry.

## Output, purity and alarms

- The plant uses PSA technology to produce medical oxygen of 93%±3 from ambient air.
- Continuous output pressure is given as 400-600 kPa (4 to 6 bar), with an alarm when the output pressure falls below 3.5 bar.
- An oxygen concentration monitor with +/- 1% accuracy, and an alarm when the concentration is lower than 90%.
- The generator section says the oxygen shall be medical grade with purity of 93±3% as defined in the European Pharmacopoeia 7.1 monograph Oxygen, 93 per cent (04/2011:2455).

## Impurity limits listed for the generator

| Impurity (at STP) | Limit in the specification |
|---|---|
| Carbon monoxide | at most 5 ppm v/v |
| Carbon dioxide | at most 300 ppm v/v |
| Water | at most 67 ppm v/v |
| Oil | at most 0.1 mg/m3 |
| Nitrogen monoxide and nitrogen dioxide | at most 2 ppm v/v |
| Sulphur dioxide | at most 1 ppm v/v |

## Feed air and local standards

The feed air compressor may be oil-free, filtered oil-injected or oil-lubricated, with a minimum of 750 kPa (7.5 bar). The compressor section says it compresses air to 5 to 10 bar to feed the PSA unit and that it must comply with ISO 8573-1:2010. For the air supplied to the PSA intake, the generator section refers to KS 2170-2:2008 (Kenya's Part 2 on medical air) or an equivalent standard. For how compressor choice interacts with a tender, see [does a PSA plant need oil-free compressors](https://library.ultracontrolo.com/answers/does-a-psa-plant-need-oil-free-compressors/).

## Power, climate and service

- Power supply: three-phase 415 VAC 50 Hz, or single-phase 240 VAC 50 Hz, with a UPS for the PSA control unit.
- The compressor section says it is suitable for use in ambient temperatures up to +45°C, and the generator section asks for continuous operation in ambient temperature of at least 5 to 45 °C, with relative humidity of at least 15 to 90% non-condensing.
- Plant life-span of a minimum of 10 years, guaranteed by a letter from the manufacturer.
- A 3-year spare parts kit as per the recommended preventive maintenance program, listed with part numbers and descriptions.
- On-site training for installation, use and maintenance is described as preferable, together with remote support.
- Easy to install: preassembled and skid-mounted, or containerized.

## Where it applies

The generator section lists levels of use: sub-county hospital (Level 4), county referral (Level 5) and national referral hospital (Level 6). The specification sets requirements for a buyer's tender; it does not say who inspects or approves a plant. For the regulator's device list, see [which Kenyan body regulates medical devices](https://library.ultracontrolo.com/answers/which-kenyan-body-regulates-medical-devices-such-as-oxygen-plants/).

### What purity does the Kenyan specification require?

The generator section says the oxygen shall be medical grade with purity of 93±3%, as defined in the European Pharmacopoeia monograph Oxygen, 93 per cent.

### Does it say anything about hot climates?

The compressor section says it is suitable for use in ambient temperatures up to +45°C, and the generator section asks for continuous operation at 5 to 45 °C or more.

## Sources

- [Kenya Ministry of Health, Medical Oxygen and Oxygen Therapy Technical Specifications for Devices and Equipments, 2024 (copy hosted by stoppneumonia.org, read 19 November 2026)](https://stoppneumonia.org/wp-content/uploads/2024/05/Technical-Specifications-for-Medical-Oxygen-Oxygen-Therapy-Devices-and-Equipments.pdf)

Reviewed: 2026-11-19. Ultra Controlo engineering. https://library.ultracontrolo.com/answers/what-does-kenyas-2024-oxygen-specification-ask-of-a-psa-plant/