# What electrical supply does a Bolivian oxygen plant need?

**What electrical supply and earthing does a Bolivian hospital oxygen plant need?**

In Bolivia's 2021 Ministry of Health call for 13 oxygen generating plants, the specification asked for a three-phase supply of 380 to 400 VAC at 50 Hz, a protection and earthing system below 5 ohm, and an electrical protection panel with resettable breakers or replaceable fuses on both live and neutral lines.

## Supply and earthing

The power section of the specification (process code MSYD/DGGH/CONV/N° 002/2021) sets three requirements for the plants: a three-phase supply of "380-400 VAC. 50 Hz", a protection and grounding system "inferior a 5 ohm" (below 5 ohm), and an electrical protection panel. This is the requirement of one Ministry of Health call and not a general Bolivian electrical code, so a hospital outside that programme should confirm the local utility and electrical rules with its own engineers.

## The protection panel

The panel is described as protection against phase loss, electric discharges, transients, overvoltage, short circuits and overloads, and other phenomena that could damage the plant and the cylinder-filling system. The sizing of the protections is to be defined according to the capacity of the equipment. The electrical system includes protection by resettable circuit breakers or replaceable fuses, installed on both the live line and the neutral line.

## Related requirements in the same document

- Each bidder states the individual electrical consumption of each item of equipment and the total consumption of the plant, together with dimensions and weight.
- The supplier prepares the technical requirements for civil works or electrical complements needed for the plant to operate, as part of pre-installation.
- Installation is included in the offer, covering the electrical installation and interconnection.
- Plants are to be designed for extreme conditions, adapted to the ground, humidity, dust and high or low temperatures of the final site, and bidders take the environmental conditions of each site into account after inspection.

The oxygen purity, monitoring and alarm requirements of the same call are in [What do Bolivian specifications require for oxygen plants?](https://library.ultracontrolo.com/answers/what-do-bolivian-specifications-require-for-oxygen-plants/). The effect of site altitude on a PSA plant is covered in How does altitude affect a PSA plant?. The specification was read on 19 November 2026.

### What earthing resistance did the Bolivian call ask for?

The specification asked for a protection and earthing system below 5 ohm.

### Does the specification say what voltage and frequency the plants use?

Yes. It asks for a three-phase supply of 380 to 400 VAC at 50 Hz.