# What a maintenance record for each device should contain

**What should a hospital's maintenance record for each piece of equipment contain, and must it be kept on a computer?**

WHO's Medical equipment maintenance programme overview (2011) says the record for each device should include identifying data such as a brief description, manufacturer, model, serial number and location, and that it is helpful to add the time and expense of scheduled and unscheduled maintenance; WHO says paper records are still useful.

WHO's 2011 overview, part of its Medical device technical series, says the data on time and expense are typically contained in work order records, which document every maintenance task performed on the device. As a result, WHO says, the inventory database will contain the entire technical and financial history for each device in the inventory.

## Why WHO stresses documentation

WHO writes that, from a regulatory point of view, work that was not documented is work that was never performed, and that it is invaluable to have thorough documentation of all service events which have taken place over the life of the equipment. Keeping it in a [CMMS](https://library.ultracontrolo.com/glossary/cmms/) for electronic retrieval is ideal in WHO's view, but simply having the information available in a paper format is still useful.

## Fields in a computerized record

WHO's separate guide to computerized maintenance management systems (2011) gives examples of table fields. The equipment model table uses a model number, a serial number and a parts list; the maintenance table uses an inventory number, a work order number, a service provider, a service engineer code and a fault code and name.

## Identification and labels

WHO says it is good practice to label each piece of medical equipment with a unique identification number, which users can quote to the maintenance department so there is no confusion about which equipment is being reported. After inspection and preventive maintenance, a label indicating the date the work was done and the procedure that was performed should be applied to the equipment.

These are WHO's recommendations for medical equipment in general. The records a hospital keeps for its gas systems under a national or local rule are a separate question: see what medical gas records a hospital must keep.

### Does WHO say maintenance records must be on a computer?

No. WHO's 2011 overview says a CMMS is ideal for electronic retrieval, but that simply having the information available in a paper format is still useful.

### Why label each device with a unique number?

WHO's 2011 overview says users can use the number to tell the maintenance department which specific piece of equipment is being reported, so there is no confusion.

## Sources

- [WHO, Medical equipment maintenance programme overview (WHO Medical device technical series, 2011), sections 5.3.5 and 5.3.7](https://iris.who.int/server/api/core/bitstreams/e5a3afa2-3eb0-41a0-820f-edb2032c41bb/content)
- [WHO, Computerized maintenance management system (WHO Medical device technical series, 2011), Table 1](https://iris.who.int/server/api/core/bitstreams/75ce2892-6c40-4f32-8277-64787ae342bf/content)

Reviewed: 2026-11-19. Ultra Controlo engineering. https://library.ultracontrolo.com/answers/what-should-a-maintenance-record-for-each-device-contain/