# What should a PSA oxygen plant tender specify?

**What should the technical specification of a PSA oxygen plant tender include?**

The 2020 WHO interim specification for PSA oxygen plants (superseded in 2024), which UNDP's 2025 requirements cite as the minimum, covered flow, purity and pressure, alarms and monitoring, the air compressor and treatment, power supply, site conditions, spare parts, documents, warranty and life span, standards, certificates, commissioning and training.

## Checklist from the published specifications

The table follows the 2020 WHO interim specification for PSA oxygen plants, adding where UNDP's REQ-OXY-001 (effective 1 January 2025) asks for more.

| Heading | What to specify | Source |
|---|---|---|
| Output | Medical oxygen 93% ±3 from ambient air; continuous flow covering 100% of the oxygen demand | WHO (2020) |
| Pressure | Continuous 300 to 600 kPa (3 to 6 bar); alarm below 3 bar | WHO (2020) |
| Monitoring and alarms | Oxygen monitor with ±1% accuracy; display of concentration, production trend, pressure, maintenance need and running hours; alarms for oxygen below 90%, high temperature, low or high pressure, power and system failure, reserve source active and dryer dew point above 3 °C | WHO (2020) |
| Feed air and treatment | Oil-free, or filtered oil-injected or oil-lubricated rotary screw, compressor of at least 750 kPa; external air dryer; pre-filter above 5 micron, 0.1 micron coalescing filter and carbon filter; receiver with bacterial outlet filter | WHO (2020) |
| Power and site | Local frequency, voltage and plug type; operation at 10 to 40 °C, 15 to 95% relative humidity and 0 to 1,000 m as a minimum; electrical requirements chart with starting current and load curves | WHO (2020); UNDP (2025) |
| Configuration | Skid-mounted, containerised or built on site; primary, secondary or backup role; connections to the filling ramp, pipeline and automatic changeover | WHO (2020); UNDP (2025) |
| Spare parts | 3-year kit listed with part numbers, descriptions and unit cost; inlet and outlet bacteria filters for 3 years | WHO (2020); UNDP (2025) |
| Documents | Manufacturer's life span letter; certificate of quality, calibration and inspection; user and service manuals; local service agent contacts; P&ID | WHO (2020); UNDP (2025) |
| Standards and certificates | ISO 7396-1, ISO 8573-1, -2 and -4, ISO 5011, ISO 21969; vessels to PED or ASME VIII, cleaned to ISO 15001 or ASTM G93; certified quality management system, for example ISO 13485; free sales certificate such as FDA or CE | WHO (2020) |
| Warranty and life span | Life span of at least 10 years guaranteed by the manufacturer; warranty of 48 months (WHO) or at least 24 months from commissioning (UNDP) | WHO (2020); UNDP (2025) |
| Commissioning and training | Verify purity, pressure, analyser, alarms and switch to the secondary supply; third-party verification of the installation; UNDP adds performance tests of at least three days of continuous operation; training on cleaning, testing and maintenance frequency | WHO (2020); UNDP (2025) |

Certificates are covered in more detail in what certificates a PSA supplier should provide.

## Decisions the specification has to make

- Purity band: ISO 7396-1:2016 states that where all sources deliver oxygen 93, the concentration may vary between 90% and 96%. A Global Fund briefing note (3 April 2024) gives the example of equipment producing 20 Nm³/h at 93% (±3%) but 18 Nm³/h at 95% (±1%).
- Technology: a Global Fund briefing note (4 March 2025) says the most important factor is the technology already common and functional in the region, and records reports of low purity with VPSA. UNDP notes that VSA and VPSA need oil-free boosters and specialised tools.
- Service: UNDP says a service level agreement should set out corrective maintenance terms, including response times and spare parts lead times.
- Editions: the WHO document was issued on 8 June 2020 as interim guidance, said it would expire two years after publication and was superseded by WHO's technical specifications of 31 December 2024, although WHO still hosts it. It cited NFPA 99 (2018), whose current edition is 2027, and ISO 7396-1:2016, whose fourth edition is under publication.

## Ultra Controlo plants

The ULTRAOX SC and ST datasheets state an oxygen purity of 93% ± 3%, and the ULTRAOX datasheet gives an outlet pressure of 6.0 bar. On ULTRAOX SC, purity is measured by one or two zirconium oxide analysers, and the equipment is parameterised and tested in the factory. ULTRAOX plants can be connected to a diesel-electric generator.

### Is the WHO PSA specification still current?

No. It was issued on 8 June 2020 as interim guidance, said it would expire two years after publication, and WHO's Technical specifications for health facility based medical oxygen systems (31 December 2024) supersede it, although WHO still hosts it and UNDP's 2025 requirements still cite it as the minimum. Check each standard it cited against its current edition.

### Should a tender ask for 93% or 95% purity?

The 2020 WHO specification used 93% ±3. A Global Fund note (2024) shows the trade-off: equipment giving 20 Nm³/h at 93% (±3%) gives 18 Nm³/h at 95% (±1%).

### How long should the warranty be?

The 2020 WHO interim specification asked for 48 months with an option to extend; UNDP's 2025 requirements ask for at least 24 months from commissioning, confirmed by the manufacturer.

## Sources

- [WHO, Technical specifications for PSA oxygen plants, interim guidance (8 June 2020)](https://iris.who.int/server/api/core/bitstreams/d53996d5-380c-44ae-9fe8-9e529d3270cd/content)
- [WHO, publication page for WHO/2019-nCoV/PSA_Specifications/2020.1](https://www.who.int/publications/i/item/WHO-2019-nCoV-PSA_Specifications-2020.1)
- [UNDP, REQ-OXY-001 v01, Technical and regulatory requirements for oxygen plants (effective 1 January 2025)](https://healthimplementation.undp.org/wp-content/uploads/2025/01/REQ-OXY-001-v01-UNDP-TECHNICAL-REGULATORY-REQUIREMENTS-FOR-OXYGEN-PLANTS.pdf)
- [Global Fund / Build Health International, Briefing Note: PSA Plant Oxygen Purity (3 April 2024)](https://resources.theglobalfund.org/media/lenpqaoa/cr_c19rm-psa-oxygen-purity_technical-briefing-note_en.pdf)
- [Global Fund / Build Health International, Briefing Note: PSA, VPSA and VSA (4 March 2025)](https://resources.theglobalfund.org/media/dp3clsrr/cr_c19rm-psa-vpsa-vsa_technical-briefing-note_en.pdf)
- [ISO 7396-1:2016, iso.org](https://www.iso.org/standard/60061.html)
- [ISO 7396-1, 4th edition project, iso.org](https://www.iso.org/standard/87169.html)
- [NFPA, NFPA 99 code development page](https://www.nfpa.org/codes-and-standards/nfpa-99-standard-development/99)
- [ULTRAOX SC datasheet, p. 2 and p. 3; 2026 edition (oxygen purity)](https://www.ultracontrolo.com/products/ultraox)
- [ULTRAOX ST datasheet, 2026 edition (oxygen purity)](https://www.ultracontrolo.com/products/ultraox-st)
- [ULTRAOX datasheet 2022, p. 5 and p. 9](https://www.ultracontrolo.com/ultraox)
- [WHO, Technical specifications for health facility based medical oxygen systems (31 December 2024), note on the oxygen generator plant specification (supersedes the 2020 interim PSA specifications)](https://iris.who.int/server/api/core/bitstreams/ff437601-add3-44c8-8149-b85f0cdb2b52/content)

Reviewed: 2026-10-22. Ultra Controlo engineering. https://library.ultracontrolo.com/answers/what-should-a-psa-tender-specification-include/