# Ph. Eur. vs USP Oxygen 93: are they the same?

Ph. Eur. monograph 2455 and the USP monograph Oxygen 93 Percent set the same oxygen range, 90.0% to 96.0%, for oxygen made from air by adsorption, but they are separate texts with their own tests, and each applies where its pharmacopoeia is the legal reference.

Both monographs cover oxygen produced from air by adsorption. They agree on the oxygen range and on what makes up the rest of the gas; they differ in publisher, legal reach, public detail and companion monographs.

|  | Ph. Eur. 2455 Oxygen (93 per cent) | USP Oxygen 93 Percent |
|---|---|---|
| Publisher | EDQM (European Pharmacopoeia) | United States Pharmacopeia (USP-NF) |
| Production covered | Single-stage adsorption plant | Molecular sieve process, from air |
| Oxygen content | Nominal 90.0% to 96.0% | 90.0% to 96.0% by volume |
| Rest of the gas | Argon and nitrogen | Mostly argon and nitrogen |
| Point where the quality is defined | Gas delivered to the pipeline system, not at the terminal outlet | Not stated in the public definition |
| Publicly readable | Status and revision history; limits only in the subscription text | Definition only; tests in the subscription text |
| Higher-purity monographs | 0417 Oxygen (cryogenic, minimum 99.5%) and 3098 Oxygen (98 per cent, two-stage) | Oxygen (not less than 99.0%) |

## What is the same

Both set an oxygen content of 90.0% to 96.0%, and both describe the remainder as argon and nitrogen. Both name an adsorption route: a single-stage adsorption plant in the Ph. Eur. and the molecular sieve process in the USP. The International Pharmacopoeia of WHO uses the same range for its Oxygen 93%, which it describes as produced from ambient air by PSA or VSA.

## What differs

- They are separate texts with their own tests, so compliance has to be shown against the monograph that applies.
- EDQM states that Ph. Eur. 2455 defines the gas delivered into the pipeline, not at the outlet; issue 12.3 removed the indicator tube tests and added a risk assessment for other impurities.
- The companion monographs do not match: Ph. Eur. 0417 Oxygen requires at least 99.5%, while USP Oxygen requires at least 99.0%.
- Europe also has Ph. Eur. 3098 Oxygen (98 per cent) for two-stage plants, with a nominal 98.0% varying between 96% and 100%.

## Which one applies

The applicable monograph is the one named by national law. In Portugal, Decree-Law 176/2006 (INFARMED consolidated text, version of 26 December 2023) requires medical gases to meet the Portuguese pharmacopoeia or, failing that, the European pharmacopoeia or that of another Member State, and AEMPS refers on-site oxygen 93 in Spain to Ph. Eur. 2455. In England, the NHS England guidance HTM 02-01 (2026) points to British Pharmacopoeia monographs. EIGA Doc 195 notes that the European monograph 2455 may be followed where there is no local pharmacopoeia. For limits that anyone can read, the [International Pharmacopoeia](https://library.ultracontrolo.com/standards/international-pharmacopoeia-oxygen/) publishes them for Oxygen 93%: carbon dioxide at most 300 ppm, carbon monoxide at most 5 ppm, nitrogen oxides at most 2 ppm in total, sulfur dioxide at most 1 ppm, oil at most 0.1 mg/m³ and water at most 67 ppm.

ULTRAOX SC and ULTRAOX ST plants supply oxygen at 93% ± 3% purity according to their datasheets.

### Does meeting USP Oxygen 93 Percent mean meeting Ph. Eur. 2455?

Not automatically. The oxygen ranges match, but each monograph has its own tests, so compliance must be shown against the one that applies in the country of use.

### Which monograph covers oxygen from a two-stage PSA plant?

In Europe, Ph. Eur. 3098 Oxygen (98 per cent), with a nominal 98.0% varying between 96% and 100%; the International Pharmacopoeia lists Oxygen 98% at not less than 96.0%.

## Sources

- [EDQM Knowledge database, monograph 2455 Oxygen (93 per cent)](https://extranet.edqm.eu/4DLink1/4DCGI/Web_View/mono/2455)
- [EDQM, monograph 2455 revision history](https://extranet.edqm.eu/4DLink1/Web_mono_hist?2455)
- [EDQM, monograph 3098 revision history](https://extranet.edqm.eu/4DLink1/Web_mono_hist?3098)
- [EDQM news on oxygen monographs 0417, 2455 and 3098](https://www.edqm.eu/en/newslist/-/asset_publisher/JaJ4OiL8a8pm/content/id/93758)
- [USP-NF monograph Oxygen 93 Percent](https://doi.usp.org/USPNF/USPNF_M59560_02_01.html)
- [USP-NF monograph Oxygen](https://doi.usp.org/USPNF/USPNF_M59550_02_01.html)
- [WHO, The International Pharmacopoeia, 13th edition (2025): Medicinal oxygen](https://digicollections.net/phint/202502/index.html#d/b.6.1.226)
- [WHO, Foundations of medical oxygen systems (17 February 2023)](https://iris.who.int/server/api/core/bitstreams/4ad6f5fd-089e-4142-8439-b64e98d816d8/content)
- [Decreto-Lei n.º 176/2006, Article 149 (INFARMED consolidated text, version of 26 December 2023)](https://www.infarmed.pt/documents/15786/1068535/035-E_DL_176_2006_12ALT/d2ae048e-547e-4c5c-873e-b41004b9027f)
- [AEMPS, note on medicinal oxygen and oxygen from hospital concentrators (18 February 2025)](https://www.aemps.gob.es/informa/la-aemps-recuerda-las-diferencias-existentes-entre-el-oxigeno-considerado-medicamento-y-el-obtenido-a-traves-de-equipos-concentradores-de-uso-hospitalario/)
- [HTM 02-01 Part A (NHS England, 2026)](https://www.england.nhs.uk/wp-content/uploads/2016/05/health-technical-memorandum-02-01-medical-gas-pipeline-systems-part-a-design-installation-validation-verification-aug-2026.pdf)
- [AIGA 113/20 (adopted from EIGA Doc 195), On site generation of oxygen 93% for medicinal use](https://asiaiga.org/uploaded_docs/en_AIGA_113_20_Generation_of_Oxygen_93_for_Medicinal_Use.pdf)
- [ULTRAOX SC and ULTRAOX ST datasheets, 2026 editions (oxygen purity)](https://www.ultracontrolo.com/products/ultraox)

Reviewed: 2026-10-22. Ultra Controlo engineering. https://library.ultracontrolo.com/compare/ph-eur-vs-usp-oxygen-93/