# Preventive vs corrective maintenance of medical gas plants

**What is the difference between preventive and corrective maintenance of medical gas and oxygen plants?**

In WHO's Medical equipment maintenance programme overview (2011), preventive maintenance is scheduled activity to extend the life of a device and prevent failure, while corrective maintenance is the process used to restore its physical integrity, safety and/or performance after a failure.

## How the sources describe each

| Aspect | Preventive maintenance | Corrective maintenance |
|---|---|---|
| WHO (2011) | Scheduled activities performed to extend the life of a device and prevent failure, for example calibration, part replacement, lubrication and cleaning | A process used to restore the device after a failure; WHO treats corrective maintenance, unscheduled maintenance and repair as equivalent terms |
| PSA plants (PATH and CHAI, 2021) | Can include cleaning or replacing filters, draining water or oil traps, replacing lubricants and calibrating pressure and oxygen sensors | May require rebuilding a valve assembly or compressor after a failure; sieve beds must be recharged periodically if oxygen output or purity falls |
| Medical gas pipelines (HTM 02-01 Part B, 2026, NHS England) | 11.1: all medical gas pipeline systems should be under a planned preventive maintenance schedule, under the responsibility of the Authorised Person (MGPS) | 11.83 and 11.84: the contractor should provide a call-out service for breakdowns between planned visits, available 24 hours per day, 365/6 days per year |
| Oxygen plant requirements (UNDP REQ-OXY-001, 2025) | 7.5: the end user, with the supplier, should implement a preventive maintenance schedule according to the manufacturer's recommendations | 6.5: an SLA, if one exists, should outline corrective maintenance terms, including estimated response times |
| Service contract (Global Fund and BHI, 2024) | Service frequency by hours of operation or calendar dates, or both | Whether part replacement, freight, labour, accommodation and travel are included in emergency repairs |

## How they are contracted

UNDP says the maintenance services must include a warranty covering a minimum of two years, encompassing preventative and corrective maintenance. PATH and CHAI (2021) say a buyer can contract with the manufacturer for maintenance that requires highly trained technicians and engineers, and that such an agreement may not always be feasible; in that case buyers should evaluate whether trained technicians or qualified third parties are available. See [service level agreement](https://library.ultracontrolo.com/glossary/service-level-agreement/) for what the agreement can specify.

## Ultra Controlo service intervals

The ULTRAOX SC datasheet states that the plant has one service at 10,000 hours, the ULTRAVAC CLP datasheet one oil change every 20,000 hours, and the ULTRASEG datasheet an annual operating inspection and eventual replacement of the protective filter element. [UltraCare](https://library.ultracontrolo.com/answers/what-is-ultracare/) covers preventive maintenance, corrective service, original spare parts and emergency support.

### Does a warranty cover both kinds of maintenance?

UNDP REQ-OXY-001 (effective 1 January 2025), section 6.5, says the maintenance services must include a warranty covering a minimum of two years, encompassing preventative and corrective maintenance services, refresher training and spare parts.

## Sources

- [WHO, Medical equipment maintenance programme overview (WHO Medical device technical series, 2011), introduction and definitions](https://iris.who.int/server/api/core/bitstreams/e5a3afa2-3eb0-41a0-820f-edb2032c41bb/content)
- [PATH and CHAI, Oxygen Generation and Storage: Pressure/Vacuum Swing Adsorption Plant (July 2021)](https://media.path.org/documents/O2_generation_and_storage_PSA_VSA_v1.pdf)
- [NHS England, HTM 02-01 Part B (2026), paragraphs 11.1, 11.83 and 11.84](https://www.england.nhs.uk/wp-content/uploads/2016/05/health-technical-memorandum-02-01-medical-gas-pipeline-systems-part-b-operational-management-aug-26.pdf)
- [UNDP, REQ-OXY-001 v01 (effective 1 January 2025), sections 6.5 and 7.5](https://healthimplementation.undp.org/wp-content/uploads/2025/01/REQ-OXY-001-v01-UNDP-TECHNICAL-REGULATORY-REQUIREMENTS-FOR-OXYGEN-PLANTS.pdf)
- [Global Fund / Build Health International, Overview: Oxygen Plant Service Contract Guidelines (20 May 2024)](https://resources.theglobalfund.org/media/aeddhtsy/cr_c19rm-oxygen-plant-service-contract_overview_en.pdf)
- [ULTRAOX SC datasheet, p. 2](https://www.ultracontrolo.com/products/ultraox)
- [ULTRAVAC CLP datasheet, p. 2](https://www.ultracontrolo.com/products/ultravac-clp)
- [ULTRASEG datasheet, p. 3](https://www.ultracontrolo.com/products/ultraseg)
- Ultra Controlo, Industrial 2026 catalogue, p. 22 (UltraCare)

Reviewed: 2026-11-18. Ultra Controlo engineering. https://library.ultracontrolo.com/compare/preventive-vs-corrective-maintenance/