# Double-stage pipeline distribution

In AIGA 113/20, a double-stage pipeline distribution system is one in which gas is initially distributed from the supply system at a pressure higher than the nominal distribution pressure and then reduced to it by line pressure regulators.

AIGA 113/20, adopted from EIGA Doc 195, says a booster compressor could be required to ensure that the supply gas is at the nominal pressure of the pipeline according to EN ISO 7396-1, usually only where a double-stage system is used, with a nominal pressure of 8 to 9 bar(g).

Under HTM 02-01 (2026), the NHS England guidance, medical gases may be distributed at a higher pressure than the nominal pressure at terminal units. The maximum working pressure should then not exceed 1100 kPa and a pressure-reducing station (PRS) should be installed. Simplex systems should not be used, and the PRS should include duplex pressure-regulating valves.

Portugal's ACSS ET 03/2006 (2024 revision) gives a primary network pressure of 800 kPa ± 100 for oxygen and 480 kPa + 100 in the secondary oxygen network, and says each service should be supplied by at least two second-reduction sets in bypass. These are Portuguese values, not a general rule.

The ULTRALOBO LPT datasheet says Ultra Controlo developed its low-pressure oxygen compression systems to meet the needs of hospitals without altering existing medical oxygen networks or compromising approved local standards. See how a hospital can use PSA oxygen without changing its network.

## Sources

- [AIGA 113/20 (adopted from EIGA Doc 195), On site generation of oxygen 93% for medicinal use, 3.2 and 4.3.1](https://asiaiga.org/uploaded_docs/en_AIGA_113_20_Generation_of_Oxygen_93_for_Medicinal_Use.pdf)
- [NHS England, HTM 02-01 Part A (2026), paragraphs 3.71, 15.164 to 15.168](https://www.england.nhs.uk/wp-content/uploads/2016/05/health-technical-memorandum-02-01-medical-gas-pipeline-systems-part-a-design-installation-validation-verification-aug-2026.pdf)
- [ACSS, ET 03/2006 v.2024, sections 6.1 and 7.3](https://www.acss.min-saude.pt/wp-content/uploads/2016/10/ET-03_2006_v2024-Gases-medicinais_VF.pdf)
- [ULTRALOBO LPT datasheet, p. 2](https://www.ultracontrolo.com/products/ultralobo)

Reviewed: 2026-11-18. Ultra Controlo engineering. https://library.ultracontrolo.com/glossary/double-stage-pipeline-distribution/