# Guidance on on-site generation of Oxygen 93 for medical use

EIGA Doc 195, On Site Generation of Oxygen 93% for Medicinal Use, is industrial gases industry guidance for designing, installing, operating, maintaining and managing the quality of oxygen 93 concentrator systems that supply hospital pipelines; the Asia Industrial Gases Association adopted it as AIGA 113/20.

## Scope

The guidance applies to the planning, technical design, location, installation, operation, maintenance and quality management of oxygen 93 concentrator systems used, where accepted, to supply medical gas pipelines in healthcare facilities, and to the roles and responsibilities needed to manage that supply. It is written for pressure swing adsorption (PSA); its principles may be applied to alternatives such as vacuum swing adsorption (VSA).

It does not address the regulatory requirements for supplying medicinal oxygen and oxygen 93 through the same pipeline, which are local or national matters, and it does not cover using an on-site generator to fill portable oxygen cylinders. It is written for healthcare facility managers and responsible qualified personnel, oxygen concentrator manufacturers, EIGA members and national regulators.

## Edition and status

This page is based on AIGA 113/20, Safe design and operation of on site generation of oxygen 93% for medicinal use, which states that it is adopted from EIGA Doc 195 with EIGA's permission. The EIGA original was not consulted, so check the current EIGA edition before citing its number or date.

## Main recommendations

- Three sources: following EN ISO 7396-1, a life support gas supply should have primary, secondary and reserve sources, each able to supply the entire flow including peak consumption.
- Power: concentrator units depend on electricity, so at least one of the three sources should be independent of the electrical supply, such as liquid oxygen, cylinders or cylinder bundles.
- Sizing: the oxygen concentration from a concentrator tends to fall as throughput rises, so the system should be designed for the maximum demand of the pipeline, including peaks. Argon, the main other component, can be present up to 5%.
- Mixed supply: where medicinal oxygen and oxygen 93 share a pipeline, the oxygen content at the outlet can vary between 90% and 100%, and a risk assessment should confirm that this is compatible with ventilators, anaesthesia machines and other medical devices.
- Anaesthesia: with low or minimal fresh gas flows argon can accumulate in the circle system, so multi-gas monitoring is essential when oxygen 93 is the oxygen source.
- Reserve refilling: filling high pressure reserve storage from the oxygen 93 system is possible but brings additional safety issues, including the risk of oxygen fires.
- Responsibility: the healthcare facility's qualified personnel, normally the head pharmacist or other staff appointed by management, are responsible for the quality and safety of the oxygen 93 given to patients.
- Validation: the process is validated through installation, operational and performance qualification (IQ, OQ, PQ).

## Related documents

The gas is defined by [Ph. Eur. 2455](https://library.ultracontrolo.com/standards/ph-eur-2455-oxygen-93/), which the guidance says may be followed where there is no local pharmacopoeia, and the pipeline by [ISO 7396-1](https://library.ultracontrolo.com/standards/iso-7396-1/). In Portugal, medical gases made in health institutions with medical devices fall under the [INFARMED regulation](https://library.ultracontrolo.com/standards/infarmed-medical-gas-regulation/).

### Who is responsible for oxygen 93 quality in a hospital, according to this guidance?

The healthcare facility's qualified personnel, normally the head pharmacist or other adequately qualified staff appointed by management.

### Does the guidance cover filling oxygen cylinders on site?

It does not cover filling portable oxygen cylinders from an on-site generator. It does discuss refilling high pressure reserve storage, and the extra safety issues this brings.

## Sources

- [AIGA 113/20, On site generation of oxygen 93% for medicinal use (adopted from EIGA Doc 195)](https://asiaiga.org/uploaded_docs/en_AIGA_113_20_Generation_of_Oxygen_93_for_Medicinal_Use.pdf)
- [EDQM Knowledge database, monograph 2455 Oxygen (93 per cent)](https://extranet.edqm.eu/4DLink1/4DCGI/Web_View/mono/2455)
- [ISO 7396-1:2016, iso.org](https://www.iso.org/standard/60061.html)
- [INFARMED, Deliberação n.º 1425/2025, Diário da República](https://files.diariodarepublica.pt/2s/2025/11/218000000/0013300160.pdf)

Reviewed: 2026-10-22. Ultra Controlo engineering. https://library.ultracontrolo.com/standards/eiga-oxygen-93-on-site/