# Oxygen flow converter: L/min, m³/h, m³ per day and kg

WHO's 2020 interim guidance converted oxygen flows by multiplying L/min by 60 to get litres per hour and dividing by 1,000 to get m³/h. The converter below also gives m³ per day, kilograms of oxygen at 1.428 g per litre at 0 °C and 1 atm (Open Critical Care) and the equivalent in J cylinders.

## The factors it uses

| From | To | Factor | Source |
|---|---|---|---|
| L/min | L/h | × 60 | WHO interim guidance (4 April 2020) |
| L/h | m³/h | ÷ 1,000 | WHO interim guidance (4 April 2020) |
| m³/h | m³ per day | × 24 | Open Critical Care |
| Litres of oxygen at 0 °C and 1 atm | grams | × 1.428 | Open Critical Care |
| Litres of oxygen at 22 °C | grams | × 1.322 | Open Critical Care |
| Litres per day | J cylinders per day | ÷ 6,800 | WHO and UNICEF (2019), nominal content of a J cylinder |

Open Critical Care works out the mass from the ideal gas law: one mole of O2 weighs 32 g and occupies 22.4 L at standard temperature and pressure, which it takes as 0 °C and 1 atm, giving 1.428 g per litre; at 22 °C a mole occupies 24.2 L, giving 1.322 g per litre.

## Check the units on the datasheet

WHO's Foundations of medical oxygen systems (2023) explains that Nm³/h refers to a reference condition of temperature and pressure, that capacity is sometimes expressed instead in Sm³/h, based on standard reference conditions, and that the unit used determines the actual gas volume produced. The converter treats m³/h and Nm³/h as the same number, so check which unit each datasheet uses before comparing plants.

The kilograms are for oxygen (O2). Oxygen 93 from a PSA plant contains 90.0% to 96.0% oxygen, the rest mainly argon and nitrogen, according to the International Pharmacopoeia (13th edition, 2025).

Published pairs to compare with the converter, as printed in their sources: WHO (2023) Table 2.1 lists 16 Nm³/h as 266 L/min, and the [ULTRAOX ST](https://library.ultracontrolo.com/products/ultraox-st/) datasheet lists model 1640 at 164 m³/h and 2,734 L/min.

### Is Nm³/h the same as m³/h?

Not always. WHO (2023) explains that Nm³/h refers to a reference condition of temperature and pressure and that some datasheets use Sm³/h instead, so the same number can mean a different volume of gas.

### Why show J cylinders per day?

To express a daily volume as a number of cylinders. The WHO-UNICEF table (2019) gives a J cylinder a nominal 6,800 L of oxygen; HTM 02-01 (2026) lists 6,500 L usable for an oxygen J cylinder, based on a 1400 kPa changeover, so a count based on usable contents is a little higher.

## Sources

- [WHO, Oxygen sources and distribution for COVID-19 treatment centres, interim guidance (4 April 2020), Table 1](https://www.who.int/publications/i/item/oxygen-sources-and-distribution-for-covid-19-treatment-centres)
- [WHO, Foundations of medical oxygen systems (17 February 2023), sections 2.1 and 2.5, Table 2.1](https://www.who.int/publications/i/item/WHO-2019-nCoV-Clinical-Oxygen-2023.1)
- [Open Critical Care, Oxygen FAQ: How much does gaseous oxygen (O2) weigh?](https://opencriticalcare.org/faq/how-much-does-gaseous-oxygen-o2-weigh/)
- [Open Critical Care, Oxygen FAQ: How do you calculate the conversion between a PSA plant's rate of oxygen production and cylinder volumes that can be filled?](https://opencriticalcare.org/faq/how-do-you-calculate-the-conversion-between-a-psa-plants-rate-of-oxygen-production-and-cylinder-volumes-that-can-be-filled/)
- [The International Pharmacopoeia, 13th edition (2025), Medicinal oxygen](https://digicollections.net/phint/202502/index.html#d/b.6.1.226)
- [ULTRAOX ST datasheet, p. 5 (model table)](https://www.ultracontrolo.com/products/ultraox-st)
- [WHO-UNICEF, Technical specifications and guidance for oxygen therapy devices (14 January 2019), Table 3.2](https://iris.who.int/server/api/core/bitstreams/5a996b5a-6803-486e-bccb-98efc398118d/content)
- [NHS England, HTM 02-01 Part A (2026), paragraph 5.14 and Table 13](https://www.england.nhs.uk/wp-content/uploads/2016/05/health-technical-memorandum-02-01-medical-gas-pipeline-systems-part-a-design-installation-validation-verification-aug-2026.pdf)