Is a PSA oxygen plant a medical device?

Is a hospital PSA oxygen plant regulated as a medical device, and in which class?

Short answer

A hospital PSA oxygen plant can be a medical device, with a class set by each market's rules and the manufacturer's intended purpose; in 2025 Spain's AEMPS said only serially made hospital concentrators connected to the network are medical devices. ULTRAOX plants are CE 1639, Class IIb (Directive 93/42/EEC). Check with the national authority.

Request a quote

Two things are regulated separately: the plant, which is equipment, and the oxygen it produces, which the 2025 INFARMED medical gas regulation in Portugal covers when it is made in health institutions with medical devices, and which the NHS England guidance HTM 02-01 (2026) treats as a medicine. This page deals with the plant. Who answers for the gas is covered in Does on-site oxygen production need a pharmacy licence?

How the EU sets a device class

In the EU, medical devices are regulated by Regulation (EU) 2017/745, which has applied since 26 May 2021. Article 51 divides devices into classes I, IIa, IIb and III, taking into account their intended purpose and inherent risks, and requires classification under the rules of Annex VIII. Annex VIII states that the rules are applied according to the intended purpose of the device. If the manufacturer and the notified body disagree on the class, the question goes to the competent authority of the Member State where the manufacturer has its registered place of business.

Classes listed by WHO

The WHO technical specifications for PSA oxygen plants (interim guidance, 8 June 2020, which the document said would expire two years after publication; superseded by WHO's technical specifications of 31 December 2024) listed the risk class of a PSA plant in several regulatory systems:

Regulatory systemClass listed by WHO (2020)
GHTFClass C (Rule 11)
United States (FDA)Class II
EU and AustraliaClass IIA
CanadaClass II

These are generic listings in a procurement document. The class of a given plant follows the classification rules applied to the manufacturer's intended purpose. In France, for example, the 2009 Afssaps recommendations for concentrators supplying a hospital medical gas network required a CE mark as a Class IIb medical device under Directive 93/42/EEC; check the version in force with ANSM.

When an installation is not a medical device

National authorities draw the line in different places. In Spain, AEMPS stated on 18 February 2025 that oxygen concentrators installed in hospitals are medical devices only when they are a standard, serially manufactured model connected to the hospital network; other units are treated as part of the hospital's own installation. The same note says that portable concentrators delivering oxygen directly to the patient are medical devices, and that after any major modification of a concentrator the gas produced must be certified against the European Pharmacopoeia.

What buyers usually ask for

  • PATH and the Clinton Health Access Initiative (July 2021) advise that PSA and VSA plants should have CE marking or approval from another stringent regulatory authority, and name ISO 7396-1 and US or EU pharmacopoeia compliance for 93% oxygen as relevant references.
  • The WHO 2020 interim specification asked for a manufacturer with a certified quality management system for medical devices, for example ISO 13485 or ISO 9001, and for the conformity of the installation to be verified by a certified third party.
  • The status of the gas is set separately: AEMPS says oxygen 93 generated on site in Spain is not a medicine, while Portugal applies its medical gas regulation to gases made in health institutions with medical devices.

The ULTRAOX SC datasheet states that the plant complies with ISO 7396-1:2016 and is designed and manufactured under a quality management system to ISO 13485:2016.

More questions

Does the WHO Class IIA listing apply to every PSA plant in the EU?

No. It is a generic listing in the 2020 interim specification; the class of each plant follows the classification rules applied to its intended purpose.

Who decides if the class of a device is disputed in the EU?

A dispute between the manufacturer and the notified body goes to the competent authority of the Member State where the manufacturer has its registered place of business.

Sources

Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.

Reviewed: · Ultra Controlo engineering · Markdown version