What operating conditions should an oxygen booster meet?
What continuous-operation and ambient conditions does WHO's specification set for a cylinder filling station and its booster?
Short answer
WHO's technical specifications of 31 December 2024 ask a filling station's booster to run continuously for at least 1500 hours without affecting its stated efficiency, and the station to work within specification at an ambient temperature of at least 5 to 40 °C and 15 to 95% relative humidity.
The cylinder filling station table in section 1.2 of WHO's Technical specifications for health facility based medical oxygen systems sets out these conditions for the booster compressor and the station, besides capacity and pressure:
- Continuous operation: the booster is to be capable of continuous operations (24 hours a day, 7 days a week) without affecting stated efficiency for at least 1500 hours of operations.
- Ambient range (environmental requirement for the filling station): continuous operations within specification at an ambient temperature of at least 5 to 40 °C (41 to 104 °F), with relative humidity from 15 to 95%.
- Elevation: the buyer specifies the elevation above sea level, in metres or feet.
- Outside the range: WHO says components of filling stations are very sensitive to environmental conditions, and asks the vendor, where the operating environment is out of the range, to propose accommodating measures to protect equipment and facilitate continuous operation.
Hours, display and servicing
The same table asks for a display of cumulative hours of operation, digital or analogue, and for a function switch with automatic and manual modes. WHO's Foundations of medical oxygen systems (2023) says the equipment maintenance schedule may relate to the number of hours operated, and gives as an example that air and booster compressors may need servicing after 2000 hours or 6 months of operations. The 2024 specifications add that their generic requirements should be adapted to the project needs before being considered binding.
What the ULTRAFIL datasheet lists
The ULTRAFIL datasheet (2026) lists for its booster automatic start stop from inlet pressure, high temperature and overpressure shutdowns, and a star delta starter with hour counter, on a 400 V / 50 Hz / 3 AC power supply. It does not list an ambient range, a continuous-duty figure or a cooling type, so these are points to put to the vendor in writing. For the rest of the WHO table see what a cylinder filling station tender should specify, and for upkeep see what maintenance an oxygen booster needs.
More questions
What should a vendor do if the site is outside the ambient range?
WHO (2024) asks the vendor to propose accommodating measures to protect equipment and facilitate continuous operation.
How soon may a booster need servicing?
WHO's Foundations of medical oxygen systems (2023) says air and booster compressors may need servicing after 2000 hours or 6 months of operations, following the manufacturer's specifications.
Sources
- WHO, Technical specifications for health facility based medical oxygen systems (31 December 2024), section 1.2, cylinder filling station
- WHO, Foundations of medical oxygen systems (17 February 2023), section 3.4
- ULTRAFIL datasheet, p. 2
Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.
Related
- AnswersWhat should an oxygen cylinder filling station tender specify?
- AnswersWhat site design conditions should a PSA tender specify?
- AnswersHow many hours a day should a PSA plant run?
- AnswersCan a containerised oxygen plant be moved?
- AnswersCan a hospital sell surplus oxygen to nearby facilities?
- AnswersCan a hospital with broken gas pipelines use a PSA plant?