Who is allowed to close a medical gas zone valve, and when?

Short answer

Under the NHS England guidance HTM 02-01 (2026), a planned closure of a zone valve (AVSU) is done by a Competent Person under a permit to work, witnessed by the Authorised Person. In a fire or major leak, the area's Designated Clinical Officer closes it and informs the Authorised Person.

Request a quote

A zone valve, called an area valve service unit (AVSU) in HTM 02-01, isolates the medical gases to one ward or department. Closing it cuts the supply to every outlet downstream, so the guidance separates planned closures, which follow the permit to work system, from emergency closures, which cannot wait for paperwork.

Planned closures

  • The permit to work (PTW) system applies to any action, such as closing an isolating valve, that restricts the supply.
  • The Authorised Person (MGPS) identifies the valves to isolate. The Competent Person (MGPS) closes the valves, AVSUs and line valve assemblies, and the Authorised Person witnesses it.
  • The Designated Clinical Officer (MGPS) of the department gives permission for the interruption and signs the PTW.
  • Outlets taken out of service are fitted with "Do not use" plugs, and clinical staff are told which outlets cannot be used.

Emergency closures

  • In a fire or a major medical gas leak, the Designated Clinical Officer for the area moves patients to an alternative gas supply, closes the AVSUs and informs the Authorised Person as soon as possible.
  • In areas where it is confirmed that no patients are connected, such as an outpatient department out of hours, isolation may be done by the Designated Clinical Officer or clinical lead, an Authorised Person, or suitably trained on-call estates staff out of hours. This does not apply to wards, theatres, critical care, diagnostics in active use or any area where patients may be connected.
  • Where it is unsafe for staff to enter, the fire and rescue service closes the valve after the Authorised Person or trained on-call estates staff have told them where it is and how it closes.
  • Repairing the leak, testing the local system and replacing the AVSU emergency access feature afterwards all need a PTW.

How the valve is protected

Valves outside locked plantrooms should be lockable in the open or closed position, but for valves that may be needed in an emergency, such as AVSUs, the lock should be capable of being overridden. AVSU keys are individually numbered and kept on individual pegs in a secure key cabinet. Designated Clinical Officers are trained to use an AVSU for emergency isolation.

Outside England

HTM 02-01 is NHS England guidance. Other countries follow their own codes, such as NFPA 99 in the United States, whose current edition is 2027. WHO describes zone valves as usually placed to be operated only by authorised personnel, and notes that in an emergency first responders can open the glass door of the valve box and close the valve.

More questions

Does closing a zone valve need a permit to work?

For planned work, yes: HTM 02-01 applies the permit to work to any action that restricts the supply, including closing an isolating valve. In an emergency the valve is closed first, and the repair that follows needs a permit.

Who reopens the valve after the work?

The Competent Person restores the gas service, witnessed by the Authorised Person, and the Designated Clinical Officer takes the area back into use once the gases have been tested and verified.