Oxygen and PSA

Portable or fixed oxygen purity analyser for a PSA plant

Short answer

The Global Fund and Build Health International sustainability guidance (22 August 2024) asks for oxygen purity to be recorded daily with at least two different analyzers, at a minimum a handheld one, while AIGA 113/20 says two analysers continuously monitor and record the concentrator output.

The documents below describe two kinds of reading: analysers connected to the plant's own control and alarm systems, which this page calls fixed, and handheld analyzers used for spot checks. They come from different authors and are written for different settings: the Global Fund material supports PSA plants in Project BOXER, AIGA 113/20 is an industry guide to on site generation of oxygen 93, and WHO-UNICEF (2019) is a specification for oxygen therapy devices.

DocumentWhat it asks for or describesKind of analyser
Global Fund daily maintenance checklistRecord the oxygen purity (%) from the PLC, and record it again using a handheld oxygen analyzer.PLC reading and handheld
Global Fund sustainability guidance (22 August 2024)Monitor and record oxygen purity daily with at least two different analyzers; at a minimum, measure purity daily with a handheld analyzer.Two different analyzers, handheld as the minimum
AIGA 113/20Two oxygen analysers continuously monitor and record the concentrator unit output, and shall be connected to the supply system control and alarm systems so that the main shut off valve can be automatically shut if the oxygen concentration falls outside the pharmacopoeia specification. Carbon monoxide, carbon dioxide and moisture analysers are connected in the same way.Analysers connected to the supply system
WHO-UNICEF (2019), chapter 8Focuses on portable handheld analysers that a biomedical engineering technician or health worker could use to test oxygen from a concentrator or from the low-pressure side of a cylinder or terminal unit source. For routine spot checks, alarms may not be necessary.Portable handheld

Calibration

AIGA 113/20 (section 9.4.4) says all analysers controlling and safeguarding the oxygen 93 produced shall be calibrated periodically using a certified calibration gas and in accordance with the manufacturer's instructions. WHO-UNICEF (2019) says galvanic cell analysers need to be calibrated and their sensors periodically replaced. See how often a PSA oxygen analyser should be checked.

Ultra Controlo analysers

The QuPHARM, launched by Ultra Controlo on 29 May 2026, is a wall-mounted purity analyser for medical oxygen generation systems. It measures O2, CO and CO2 concentrations simultaneously, with an optional H2O transmitter, and has programmable alarm thresholds on the O2, CO and CO2 channels. See QuPHARM.

The ULTRAOX SC datasheet lists one or two oxygen analysers with a zirconium oxide probe. See zirconia and paramagnetic oxygen analysers and how oxygen purity is measured in a PSA plant.

More questions

What does the Global Fund daily checklist record for oxygen purity?

Its data collection section asks for the oxygen purity (%) from the PLC and, separately, the oxygen purity (%) measured with a handheld oxygen analyzer.

What analysers does AIGA 113/20 list for the concentrator output?

Two oxygen analysers that continuously monitor and record the concentrator unit output, plus carbon monoxide, carbon dioxide and moisture analysers connected to the supply system control and alarm systems.

Sources

Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.

Reviewed: · Ultra Controlo engineering · Markdown version