Ultra Controlo Medical gas library

Vacuum and AGSS

Wet vs dry dental suction systems

Short answer

NHS England's HTM 02-01 (2026) classes dental vacuum systems as dry, semi-dry or wet: in dry systems an air separator removes liquids and debris before the air reaches a separate pump, while in wet systems air and liquid enter the pump to be separated; it says installations should be assessed individually.

In semi-dry systems, a similar separation takes place but the vacuum pump and the separator are combined in one device. In wet systems, a filter first removes the larger solids. HTM 02-01 does not describe the merits of each system, because installations should be assessed on an individual basis to find the most effective solution.

HTM 02-01 (2026), Part A, chapter 11Dry systemWet system
SeparationAn air separator removes dental detritus, sputa and cooling water from the airflow and passes them to a drain before the air enters the pump; separator and pump are two different devicesA filter removes the larger solids; air and liquid then enter the vacuum pump, where they are separated
Vacuum lineRelatively dry and clean, which helps minimise bacterial contamination of the line; where the separator sits close to the pumps, much of the pipework still carries detritusLiquid is also aspirated through the pipework (as in semi-dry systems)
Pipework layoutA series connection is often used, as only air is aspirated through the pumpsA star arrangement, with each chair draining back through its own pipe to the pump/separator unit, is quite common, because liquid in a series layout could cause a loss of performance at some chairs
Bacteria filterMay be feasible and is certainly desirable between the (condensate) separator and the pump inletsNot practicable upstream; where upstream filtration is not possible, a bacteria filter should be fitted in the pump exhaust line
Plant conditionsExcessive condensation can occur where pipework is exposed to low temperatures; if necessary, a condensation separator with an automatic drain should be fittedKeeping a minimum temperature of 10 °C is particularly important; freezing could cause considerable damage

Other points in HTM 02-01

  • An amalgam separator should be included in all systems.
  • Electrically driven side-channel pumps are frequently used in dental vacuum systems and may be combined with a separator unit; liquid ring pump systems typically include water-saving measures and may be fed from a reservoir or from the water mains via an air gap.
  • Medical vacuum systems should not be used to extract bodily fluids during dental procedures.

See what a dental central suction system is for how HTM 02-01 sizes these plants.

ULTRAVAC DENTAL

ULTRAVAC DENTAL is the Ultra Controlo centralised suction system for wet aspiration. It uses SCB side channel vacuum pumps with speed variation, a QuVAC controller that rotates the pumps and starts them progressively, and stainless steel vacuum stabilisation tanks. The datasheet table gives the number of chairs per model in two columns, 100% and 60%, from the V300 (1 chair) to the V18000 (60 chairs at 100%, 100 chairs at 60%).

More questions

Can a dental suction system have a bacteria filter before the pump?

HTM 02-01 (2026) says it may be feasible and is certainly desirable in dry systems to fit one between the separator and the pump inlets; in wet systems this is not practicable.

What minimum temperature does a wet dental vacuum plant need?

HTM 02-01 (2026) says maintaining a minimum temperature of 10 °C is particularly important for wet vacuum systems, because freezing could cause considerable damage.

Sources

Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.

Reviewed: · Ultra Controlo engineering · Markdown version