ULTRAGEC: specifications and models
ULTRAGEC is an emergency cabinet for critical hospital areas that keeps oxygen, medical air and vacuum available when the networks feeding the area fail.
It connects by derivation to the secondary networks upstream of the critical area and comes in 2 and 4 cylinder versions and a mobile version.
Key facts from the datasheet
- The emergency cabinet ensures permanent supply of medical gases in the most critical areas.
- It connects by derivation to secondary networks upstream of the critical areas.
- Each cabinet has a vacuum pump with a capacity of 40 m3/h.
- ULTRAGEC 2 takes 2 B50 cylinders (O2, medical air).
- ULTRAGEC 4 takes 4 B50 cylinders (O2, medical air).
- ULTRAGEC 2 has an alarm trigger through a pneumatic command.
- ULTRAGEC 4 has a QuGEC supervision monitor.
- An ULTRAGEC MOBILE version has the same features on wheels.
Models
| Electric Supply | Cylinders | Vacuum Pump | Dimensions (W x L x H) | Article | Equipment quantity: Cabinet | Equipment quantity: Oxygen | Equipment quantity: Medical Air | Equipment quantity: Vacuum Pump | Equipment quantity: Surveillance System | Designation |
|---|---|---|---|---|---|---|---|---|---|---|
| 230 V -400 V / 50 Hz | Local for 2 cylinders B50 (O2, AC) | capacity of 40m3/h | 1260 x 430 x 1820 in mm | 20230800412 | 1 | 1 | 1 | 1 | 1 | ULTRAGEC 2. 50/40 |
| Electric Supply | Cylinders | Vacuum Pump | Dimensions (L x W x H) | Article | Supervision | Equipment quantity: Cabinet | Equipment quantity: Oxygen | Equipment quantity: Medical Air | Equipment quantity: Vacuum Pump | Equipment quantity: Surveillance System | Designation |
|---|---|---|---|---|---|---|---|---|---|---|---|
| 230 V / 50 Hz | Local for 4 cylinders B50 (O2, AC) | capacity of 40m3/h | 1800 x 430 x 1820 mm | 20230800413 | QuGEC Supervision Monitor | 1 | 2 | 2 | 1 | 1 | ULTRAGEC 4. 50/40 |
Values as printed in the Ultra Controlo datasheet. Ultra Controlo engineering confirms the configuration for each project.
Sources
Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.
Related
- AnswersCan a probe fit the wrong medical gas outlet?
- AnswersAt what pressure should a medical gas area alarm trigger?
- ComparisonsCopper vs stainless steel for medical gas pipes
- AnswersDo copper medical gas pipes need earthing or isolation from supports?
- ComparisonsEN 13348 vs ASTM B819: copper tube for medical gases
- AnswersHow are empty cylinders changed on a manifold?