Ultra Controlo Medical gas library

Oxygen and PSA

ULTRAOX SC: specifications and models

ULTRAOX SC is an on-site PSA medical oxygen plant built around oil-less ULTRACECO compressors and ULTRACON concentrators, with a filling station option.

The plant is modular, runs automatically in cascade and is supplied pre-wired and factory tested.

Key facts from the datasheet

  • The ULTRAOX SC plant supplies medical oxygen at a purity of 94% (plus or minus 1%).
  • The stated purity tolerance is plus or minus 1%.
  • The basic configuration uses one or more oil-less ULTRACECO compressors.
  • The datasheet describes the ULTRACECO compressor as an oil-less air compressor of Class 0.
  • The datasheet states that the ULTRAOX SC has one service at 10,000 hours.
  • The plant has one or two oxygen analysers with a zirconium oxide probe.
  • The plant is modular and its flow capacity can be increased progressively, so no total back-up compressor is needed.
  • Compressors switch and generators start in cascade to keep a continuous oxygen supply.
  • The equipment is parameterised and tested in the factory before delivery.
  • Particles in the oxygen are below 1 µm.
  • Oil in fluid phase is listed as exempt.
  • CO2 is below 300 ppm.
  • CO is below 5 ppm.
  • SO2 is below 1 ppm.
  • NOx is below 2 ppm.
  • The working dew point is minus 46 °C.
  • The quality values are given as residual contents in initial operation conditions, with oxygen quality per the European Pharmacopoeia.

Models

MODELCAPACITY O2 per unit (l/min)CAPACITY O2 per unit (m3/h)COMPRESSOR (per unit) KWINLET PRESSURE (AC) barOUTLET PRESSURE (O2) barULTRACON
35,30,3182,27,561
8130,7772,27,562
1217,91,07247,563
3051,73,15,57,564
5076,74,67,57,565
70108,36,5137,566
90133,38137,567
100158,39,5157,568
120206,712,422,57,569
180303,318,237,57,5610
260436,726,252,57,5611
370608,336,57,5612
44072543,5757,5613
560928,355,797,57,5614

Values as printed in the Ultra Controlo datasheet. Ultra Controlo engineering confirms the configuration for each project.

Standards cited in the datasheet

  • ISO 7396-1:2016
  • ISO 13485:2016
  • European Pharmacopoeia
  • European and American Pharmacopoeia

Sources

Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.

Reviewed: · Ultra Controlo engineering · Markdown version