Medical gas library

Standards

ISO 8573-2: how oil aerosol in compressed air is measured

ISO 8573-2:2018, Compressed air: Contaminant measurement, Part 2: Oil aerosol content, is the ISO standard that specifies test methods for the sampling and quantitative analysis of liquid oil and oil aerosols that can typically be present in compressed air; oil vapour is covered by ISO 8573-5.

Edition and status

ISO 8573-2:2018 is edition 3, published in February 2018, with 45 pages, prepared by ISO/TC 118/SC 4 under ICS 71.100.20. On 19 November 2026 ISO lists it as published at stage 90.60, close of review. ISO lists ISO 8573-2:2007, now withdrawn, as the previous edition.

The two methods

ISO's abstract describes two methods. Method A is an oil collection technique using inline coalescing filters. Method B uses sampling discs in a holder, from which the collected oil is extracted with a solvent and analysed by infrared spectrometry or by gas chromatography with flame ionization detection. Method B is subdivided into two parts, which distinguish the procedures for obtaining the quantity of oil for analysis. The abstract also describes alternative oil aerosol detection with indicator type devices in Annex E.

What it leaves out

Test methods for oil vapour are excluded from this part because they are covered by ISO 8573-5. The classes of oil content themselves are set in ISO 8573-1, and Class 0 is explained in its own page.

On Ultra Controlo datasheets

The ULTRACECO SC datasheet describes compression exempt of oil or grease. It does not cite ISO 8573-2 and gives no oil aerosol test result, so this page makes no claim about the oil content of the air it delivers.

More questions

Does ISO 8573-2 cover oil vapour?

No. ISO's abstract says test methods for oil vapour are excluded because they are covered by ISO 8573-5.

What are Method A and Method B?

Method A collects oil with inline coalescing filters. Method B collects it on sampling discs in a holder, extracts it with a solvent and analyses it by infrared spectrometry or gas chromatography.

Sources

Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.

Reviewed: · Ultra Controlo engineering · Markdown version