Medical gas library

Oxygen and PSA

Can a PSA or medical air plant be installed outdoors or on a roof?

Can a PSA oxygen plant or medical air plant be installed outdoors, on a roof or in a container yard, and what protection does it need?

Short answer

Global Fund site guidance (May 2025) lists a containerised PSA plant on a concrete slab, with a canopy strongly recommended, as one of three ways to house a plant, and UNDP (2025) asks for housing that protects it from precipitation, wind and dust; neither document sets rules for a roof installation.

The Global Fund overview of oxygen plant site considerations (14 May 2025), by Build Health International, covers three ways to house a PSA plant: an existing room or building retrofitted to the vendor's requirements, a new plant building, and a containerised plant, which requires a concrete slab of at least the vendor's minimum size and a canopy, strongly recommended and in some cases required, to protect the container and its operators.

Outdoors or in a container yard

  • Weather: the container checklist asks whether the container will be exposed to blowing dust, dirt, sand, rain, sleet or snow, which informs the canopy design, as extended roof projections or other opening protection may be needed.
  • Heat: the checklist notes that PSA plant equipment can only operate in a specific temperature range, which varies by manufacturer but often does not exceed 40 °C, and asks whether the space inside a container under a canopy is likely to exceed 40 °C, in which case cooling systems need to be considered.
  • Flooding: on a site prone to flooding, the checklist says equipment will need to be raised to an appropriate height; the 2024 form asks at what height the slab should be laid, taking the flood risk into account.
  • Ground and access: it says sloping sites need extra earthwork to level the concrete pad, and asks whether a crane of sufficient capacity can reach the site on roads free from low power lines and other height restrictions.
  • Security: it notes that containerised plants may have equipment on their exterior, and that access to that space should be restricted.
  • Pollution and fire: it says smoke and exhaust sources, such as diesel generators and vehicle exhausts, and open flames must be more than 10 meters from the PSA plant.

UNDP's requirements (REQ-OXY-001, effective 1 January 2025) add that the oxygen production system must be housed in a structure that protects it from precipitation, wind and dust and reduces operational noise, that enhanced ventilation is needed if metallic roofing and corrugated sheets cause overheating, and that the nitrogen discharge duct should be vented at least 3 meters from doors or windows and face downward. AIGA 113/20, adopted from EIGA Doc 195, asks for outdoor pressure relief devices to have discharge outlets protected from weather and water freezing, and recommends insulating process, instrument and compressed air lines exposed to cold ambient conditions against freezing.

On a roof

None of these documents gives specific rules for a roof installation. For an existing structure, the Global Fund checklist asks whether it is structurally sound, since it must support large, heavy mechanical equipment, and whether a forklift or pallet jack can bring the equipment in. UNDP's site evaluation checklist (Annex II, adapted from a 2023 WHO site evaluation annex) asks whether the concrete slab load resistance is calculated for a minimum of 1000 kg/m² and whether the housing is in a floodable or seismic area. AIGA 113/20 asks for guard rails, platform gates and ladder enclosures to prevent falls from elevated locations.

Medical air plants

For medical air, HTM 02-01 Part A (2026), the NHS England guidance, says that medical air, vacuum and AGSS plant should be housed in weatherproof enclosures, preferably as part of the main building in purpose-built plantrooms, located with regard to air intake and exhaust requirements, access for servicing and major component replacements, and the routes of the medical gas pipeline.

Ultra Controlo

The ULTRAOX CO oxygen plant is built inside a sea container with all equipment included, according to the ULTRAOX datasheet (2022), and the ULTRACON datasheet lists container and skid-mounted solutions; see what a containerised oxygen plant is.

More questions

Does a containerised PSA plant need a canopy?

Global Fund site guidance (2025) says a canopy is strongly recommended, and in some cases required, to protect the container and its operators.

Can medical air compressors be installed outside?

NHS England's HTM 02-01 (2026) says medical air plant should be housed in weatherproof enclosures, preferably as part of the main building in purpose-built plantrooms.

Sources

Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.

Reviewed: · Ultra Controlo engineering · Markdown version