Short answer
According to the 2026 ULTRAOX ST datasheet, the compressor of each ULTRAOX model is sized for three altitude bands, 0 to 1,000 m, 1,000 to 2,000 m and 2,000 to 3,000 m, while the oxygen flow of the model stays the same in all three.
Altitude is one of the site conditions a plant has to be designed for. WHO (2023) says that during design the site-specific environmental considerations, such as elevation above sea level, temperature, humidity and dust, should be addressed, and UNDP (2025) says site-specific factors like elevation, temperature, humidity and dust should be considered during design. Open Critical Care, in an answer by Assist International, says plants are rated at 1000 ft above sea level or lower and that much of Sub Saharan Africa is above this, so the plants have to be derated significantly.
What the ULTRAOX ST table gives
The 2026 ULTRAOX ST datasheet lists the compressor of each model for three altitude bands: 0 to 1,000 m, 1,000 to 2,000 m and 2,000 to 3,000 m. The oxygen flow of the model is the same in all three.
| Model | Oxygen, Nm³/h | ULTRASCREW at 0 to 1,000 m | ULTRASCREW at 1,000 to 2,000 m | ULTRASCREW at 2,000 to 3,000 m |
|---|---|---|---|---|
| ULTRAOX 30 | 3.10 | ULTRASCREW 5 | ULTRASCREW 7 | ULTRASCREW 7 |
| ULTRAOX 50 | 4.60 | ULTRASCREW 7 | ULTRASCREW 11 | ULTRASCREW 11 |
| ULTRAOX 70 | 6.50 | ULTRASCREW 11 | ULTRASCREW 15 | ULTRASCREW 15 |
| ULTRAOX 90 | 8.00 | ULTRASCREW 15 | ULTRASCREW 18 | ULTRASCREW 18 |
| ULTRAOX 100 | 9.50 | ULTRASCREW 18 | ULTRASCREW 18 | ULTRASCREW 18 |
| ULTRAOX 120 | 12.4 | ULTRASCREW 18 | ULTRASCREW 22 | ULTRASCREW 30 |
| ULTRAOX 180 | 18.2 | ULTRASCREW 30 | ULTRASCREW 37 | ULTRASCREW 37 |
| ULTRAOX 260 | 26.2 | ULTRASCREW 37 | ULTRASCREW 45 | ULTRASCREW 55 |
| ULTRAOX 370 | 36.5 | ULTRASCREW 55 | ULTRASCREW 75 | ULTRASCREW 75 |
| ULTRAOX 440 | 43.5 | ULTRASCREW 75 | ULTRASCREW 75 | ULTRASCREW 75 |
| ULTRAOX 520 | 52.4 | ULTRASCREW 75 | ULTRASCREW 90 | ULTRASCREW 110 |
| ULTRAOX 560 | 55.7 | ULTRASCREW 75 | ULTRASCREW 110 | ULTRASCREW 110 |
| ULTRAOX 730 | 73.0 | ULTRASCREW 110 | ULTRASCREW 132 | ULTRASCREW 160 |
| ULTRAOX 820 | 82.7 | ULTRASCREW 132 | ULTRASCREW 160 | ULTRASCREW 160 |
| ULTRAOX 1120 | 111.4 | ULTRASCREW 160 | ULTRASCREW 200 | ULTRASCREW 200 |
| ULTRAOX 1640 | 165.3 | ULTRASCREW 250 | ULTRASCREW 315 | 2 × ULTRASCREW 160 |
Simplex layout. The compressor is an ULTRASCREW VSD. At 2,000 to 3,000 m, ULTRAOX 1640 uses 2 × ULTRASCREW 160.
For some models the compressor is the same in more than one band: ULTRAOX 100 is listed with ULTRASCREW 18 and ULTRAOX 440 with ULTRASCREW 75 in all three. The ULTRAOX ST datasheet states that the power and pressure of the compressor can vary with the service conditions required for the facility.
ULTRAOX SC
The ULTRAOX SC datasheet says that factory integration and site-specific sizing take local climate and installation conditions into account, in three altitude bands up to 3,000 m, and its table lists the same oxygen capacity in all three bands. The table does not list the compressor of each model; its notes say compressor data for the other models is available on request. At 2,000 to 3,000 m the notes list 2 compressors for the ULTRAOX 1640.
What to tell the manufacturer
UNDP's site evaluation form asks for the altitude in metres above sea level, and Open Critical Care says the manufacturer will need to know the location where the plant is to be installed. See what information is needed for an oxygen plant quote and the general answer how altitude affects a PSA plant.
More questions
Does altitude change the oxygen flow of an ULTRAOX model?
In the 2026 tables each model lists the same oxygen flow in the three altitude bands. What changes in the ULTRAOX ST is the ULTRASCREW compressor.
Which compressor does the ULTRAOX 120 use at 2,000 to 3,000 m?
The 2026 ULTRAOX ST table lists ULTRASCREW 30 for the ULTRAOX 120 at 2,000 to 3,000 m, against ULTRASCREW 18 at 0 to 1,000 m and ULTRASCREW 22 at 1,000 to 2,000 m.
Sources
- ULTRAOX ST datasheet, 2026 edition (model table, altitude bands)
- ULTRAOX SC datasheet, 2026 edition (model table and notes)
- WHO, Foundations of medical oxygen systems (WHO/2019-nCoV/Clinical/Oxygen/2023.1, 17 February 2023), section 2.2.1
- UNDP, REQ-OXY-001 v01, Technical and regulatory requirements for oxygen plants (effective 1 January 2025), section 6.2 and Annex II
- Open Critical Care, Oxygen FAQ (answer by Assist International): How do you calculate the conversion between a PSA plant's rate of oxygen production and cylinder volumes that can be filled?
Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.