Medical gas library

Oxygen and PSA

ULTRAOX models compared: 3.10 to 165.3 Nm³/h

How do the 16 ULTRAOX models compare, from ULTRAOX 30 to ULTRAOX 1640?

Short answer

According to the 2026 ULTRAOX SC and ULTRAOX ST datasheets, there are 16 ULTRAOX models, from ULTRAOX 30 at 3.10 Nm³/h of oxygen to ULTRAOX 1640 at 165.3 Nm³/h, with the same models and flows in both versions and a purity of 93% ± 3%.

Each model in the 2026 ULTRAOX SC and ULTRAOX ST datasheets is listed with its oxygen flow in Nm³/h, the normal cubic metres per hour explained in normal cubic metre. ULTRAOX 1640, for example, is listed at 165.3 Nm³/h and ULTRAOX 30 at 3.10 Nm³/h. To convert to litres per minute, use the oxygen flow unit converter.

The 16 models at a glance

ModelOxygen, Nm³/hMax. daily filling capacity, cylinders (50 L at 150 bar)Oxygen generatorULTRASCREW in ULTRAOX ST at 0 to 1,000 m / 1,000 to 2,000 m / 2,000 to 3,000 m
ULTRAOX 303.1010ULTRACON 45 / 7 / 7
ULTRAOX 504.6014ULTRACON 57 / 11 / 11
ULTRAOX 706.5020ULTRACON 611 / 15 / 15
ULTRAOX 908.0025ULTRACON 715 / 18 / 18
ULTRAOX 1009.5030ULTRACON 818 / 18 / 18
ULTRAOX 12012.440ULTRACON 918 / 22 / 30
ULTRAOX 18018.258ULTRACON 1030 / 37 / 37
ULTRAOX 26026.284ULTRACON 1137 / 45 / 55
ULTRAOX 37036.5116ULTRACON 1255 / 75 / 75
ULTRAOX 44043.5139ULTRACON 1375 / 75 / 75
ULTRAOX 52052.4168ULTRACON 11 × 275 / 90 / 110
ULTRAOX 56055.7178ULTRACON 1475 / 110 / 110
ULTRAOX 73073.0232ULTRACON 12 × 2110 / 132 / 160
ULTRAOX 82082.7264ULTRACON 15132 / 160 / 160
ULTRAOX 1120111.4356ULTRACON 14 × 2160 / 200 / 200
ULTRAOX 1640165.3528ULTRACON 15 × 2250 / 315 / 2 × 160

Simplex layout. The oxygen flow of each model is the same in the three altitude bands of the table; in the ULTRAOX ST the ULTRASCREW compressor changes with the band.

ULTRAOX 520, 730, 1120 and 1640 are listed with two ULTRACON generators of the same model, written × 2 in the table. At 2,000 to 3,000 m the ULTRAOX ST table lists two ULTRASCREW 160 compressors for ULTRAOX 1640.

What differs between ULTRAOX SC and ULTRAOX ST

The two tables list the same 16 models, the same oxygen flows, the same ULTRACON generators and the same tank sizes. They differ in the compressor: ULTRAOX SC uses oil-less ULTRACECO compressors, and ULTRAOX ST uses ULTRASCREW screw compressors, which the ULTRASCREW datasheet describes as oil-injected rotary screw compressors.

The ULTRAOX ST components table also lists an ULTRAK activated carbon tower and an ULTRASEP oil/water separator for each model; the ULTRAOX SC components table does not list them. The full list is in what equipment comes with each ULTRAOX model.

Choosing a model

The ULTRAOX product pages describe the size match as indicative, from the datasheet, with Ultra Controlo engineers confirming the configuration for the site. For sizing from beds, see how many beds a PSA oxygen plant can supply; for the compressor at altitude, see how an ULTRAOX plant is sized for altitude. One model has its own page: ULTRAOX 1640.

More questions

What flow does the ULTRAOX 1640 produce?

The 2026 ULTRAOX SC and ULTRAOX ST datasheets list ULTRAOX 1640 at 165.3 Nm³/h of oxygen, with two ULTRACON 15 generators. The tables give flow in Nm³/h, not in litres per minute.

Do ULTRAOX SC and ULTRAOX ST have the same models?

Yes. The 2026 tables list the same 16 models and the same oxygen flow for both versions. They differ in the compressor: oil-less ULTRACECO in ULTRAOX SC and ULTRASCREW screw compressors in ULTRAOX ST.

Does the oxygen flow of a model change with altitude?

In the 2026 tables each model lists the same oxygen flow at 0 to 1,000 m, 1,000 to 2,000 m and 2,000 to 3,000 m. In the ULTRAOX ST, the ULTRASCREW compressor changes with the altitude band.

Sources

Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.

Reviewed: · Ultra Controlo engineering · Markdown version

Related