Oxygen flow converter: L/min, m³/h, m³ per day and kg

WHO's 2020 interim guidance converted oxygen flows by multiplying L/min by 60 to get litres per hour and dividing by 1,000 to get m³/h. The converter below also gives m³ per day, kilograms of oxygen at 1.428 g per litre at 0 °C and 1 atm (Open Critical Care) and the equivalent in J cylinders.

Result
L/min 
L/h 
m³/h 
m³ per day 
kg of O2 per day at 0 °C and 1 atm 
kg of O2 per day at 22 °C 
J cylinders per day (6,800 L nominal) 

The converter treats m³/h and Nm³/h as the same number. Check the reference conditions each datasheet uses.

The factors it uses

FromToFactorSource
L/minL/h× 60WHO interim guidance (4 April 2020)
L/hm³/h÷ 1,000WHO interim guidance (4 April 2020)
m³/hm³ per day× 24Open Critical Care
Litres of oxygen at 0 °C and 1 atmgrams× 1.428Open Critical Care
Litres of oxygen at 22 °Cgrams× 1.322Open Critical Care
Litres per dayJ cylinders per day÷ 6,800WHO and UNICEF (2019), nominal content of a J cylinder

Open Critical Care works out the mass from the ideal gas law: one mole of O2 weighs 32 g and occupies 22.4 L at standard temperature and pressure, which it takes as 0 °C and 1 atm, giving 1.428 g per litre; at 22 °C a mole occupies 24.2 L, giving 1.322 g per litre.

Check the units on the datasheet

WHO's Foundations of medical oxygen systems (2023) explains that Nm³/h refers to a reference condition of temperature and pressure, that capacity is sometimes expressed instead in Sm³/h, based on standard reference conditions, and that the unit used determines the actual gas volume produced. The converter treats m³/h and Nm³/h as the same number, so check which unit each datasheet uses before comparing plants.

The kilograms are for oxygen (O2). Oxygen 93 from a PSA plant contains 90.0% to 96.0% oxygen, the rest mainly argon and nitrogen, according to the International Pharmacopoeia (13th edition, 2025).

Published pairs to compare with the converter, as printed in their sources: WHO (2023) Table 2.1 lists 16 Nm³/h as 266 L/min, and the ULTRAOX ST datasheet lists model 1640 at 164 m³/h and 2,734 L/min.

More questions

Is Nm³/h the same as m³/h?

Not always. WHO (2023) explains that Nm³/h refers to a reference condition of temperature and pressure and that some datasheets use Sm³/h instead, so the same number can mean a different volume of gas.

Why show J cylinders per day?

To express a daily volume as a number of cylinders. The WHO-UNICEF table (2019) gives a J cylinder a nominal 6,800 L of oxygen; HTM 02-01 (2026) lists 6,500 L usable for an oxygen J cylinder, based on a 1400 kPa changeover, so a count based on usable contents is a little higher.

Sources

Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.

Reviewed: · Ultra Controlo engineering · Markdown version