Medical gas library

Oxygen and PSA

Oxygen cylinder duration calculator

WHO and UNICEF (2019) list nominal oxygen contents from 340 L for a D cylinder to 6,800 L for a J cylinder, and a cylinder supplies about as many minutes as its litres divided by the flow in L/min. The calculator below does that division for one cylinder or a bank of cylinders.

Result 

Nominal contents from WHO and UNICEF (2019); with a gauge reading or another cylinder, WHO's rule of thumb (bar × water capacity in litres). NHS England (2023) asks staff to check the contents display and not to let a cylinder fully empty while a patient still needs oxygen.

How the calculator works

  • Lettered sizes. For D to J it uses the nominal oxygen contents in the WHO-UNICEF Technical specifications and guidance for oxygen therapy devices (2019), all at a nominal service pressure of 137 bar.
  • Gauge reading. If you enter what the pressure gauge shows, it uses the rule of thumb in WHO's Foundations of medical oxygen systems (2023): pressure in bar multiplied by water capacity in litres approximates the gas volume in litres.
  • Other cylinders. For any other cylinder, enter its water capacity and filling pressure, and the same rule of thumb applies. WHO (2023) notes that nominal pressures vary by manufacturer, typically 137 to 150 bar.
  • Time. The litres of oxygen divided by the flow in L/min give the minutes of supply; for a bank, the litres of all the cylinders are added first.
SizeWater capacity (L)Nominal oxygen content (L)Nominal service pressure
D2.3340137 bar
E4.7680137 bar
F9.41,360137 bar
G23.63,400137 bar
J47.26,800137 bar

The WHO-UNICEF table comes from a gas supplier's data. Medical cylinder sizes are named alphabetically, and other suppliers and countries use other sizes and pressures: see what sizes medical oxygen cylinders come in.

Before relying on the result

  • Check what is left. NHS England's guidance on the safe use of oxygen cylinders (24 January 2023) asks staff to always check the cylinder contents display and estimate the approximate residual volume according to the prescribed flow rate.
  • Keep a margin. The same guidance says care should be taken that cylinders do not fully empty while patients still need oxygen, and that patients transferred on a cylinder should have enough oxygen for the transfer and any diagnostic tests.
  • Usable is less than nominal. HTM 02-01, the NHS England guidance (2026), counts usable contents when sizing manifolds: its Table 13 lists oxygen J cylinders at 6,800 L nominal and 6,500 L usable, based on a 1400 kPa changeover, and says contents information should be obtained from the gas supplier.
  • Use the real flow. Flow depends on the device: WHO's Foundations of medical oxygen systems (2023) lists 1 to 5 L/min for a nasal cannula and 10 to 60 L/min for high-flow nasal oxygen.

More questions

Does the gauge reading change the result?

Yes. With a reading, the calculator estimates the litres left from WHO's rule of thumb, the pressure in bar multiplied by the water capacity in litres, instead of the nominal content of a full cylinder.

Can it work out a cylinder manifold?

It gives the time a bank of cylinders lasts at a given flow. HTM 02-01 sizes manifolds on usable rather than total contents and says contents information should be obtained from the gas supplier.

Why is the result an estimate?

The table gives nominal contents, and NHS England asks staff to check the cylinder contents display and not to let a cylinder fully empty while a patient still needs oxygen.

Sources

Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.