Short answer
Kenya's Ministry of Health 2024 technical specification for medical oxygen devices describes a PSA plant that makes medical oxygen of 93%±3 from ambient air, at a continuous output pressure of 400-600 kPa, with a feed air compressor of at least 750 kPa.
Request a quoteBuying a PSA oxygen plant in Kenya
This answer summarises the PSA plant section of the Ministry of Health document, read on 19 November 2026 from a copy hosted at stoppneumonia.org. The PDF carries the Ministry of Health cover and the date 2024; its PSA sections are marked version 1, November 2023. I did not find the document on the Ministry's own website, so a buyer should check the current version with the Ministry.
Output, purity and alarms
- The plant uses PSA technology to produce medical oxygen of 93%±3 from ambient air.
- Continuous output pressure is given as 400-600 kPa (4 to 6 bar), with an alarm when the output pressure falls below 3.5 bar.
- An oxygen concentration monitor with +/- 1% accuracy, and an alarm when the concentration is lower than 90%.
- The generator section says the oxygen shall be medical grade with purity of 93±3% as defined in the European Pharmacopoeia 7.1 monograph Oxygen, 93 per cent (04/2011:2455).
Impurity limits listed for the generator
| Impurity (at STP) | Limit in the specification |
|---|---|
| Carbon monoxide | at most 5 ppm v/v |
| Carbon dioxide | at most 300 ppm v/v |
| Water | at most 67 ppm v/v |
| Oil | at most 0.1 mg/m3 |
| Nitrogen monoxide and nitrogen dioxide | at most 2 ppm v/v |
| Sulphur dioxide | at most 1 ppm v/v |
Feed air and local standards
The feed air compressor may be oil-free, filtered oil-injected or oil-lubricated, with a minimum of 750 kPa (7.5 bar). The compressor section says it compresses air to 5 to 10 bar to feed the PSA unit and that it must comply with ISO 8573-1:2010. For the air supplied to the PSA intake, the generator section refers to KS 2170-2:2008 (Kenya's Part 2 on medical air) or an equivalent standard. For how compressor choice interacts with a tender, see does a PSA plant need oil-free compressors.
Power, climate and service
- Power supply: three-phase 415 VAC 50 Hz, or single-phase 240 VAC 50 Hz, with a UPS for the PSA control unit.
- The compressor section says it is suitable for use in ambient temperatures up to +45°C, and the generator section asks for continuous operation in ambient temperature of at least 5 to 45 °C, with relative humidity of at least 15 to 90% non-condensing.
- Plant life-span of a minimum of 10 years, guaranteed by a letter from the manufacturer.
- A 3-year spare parts kit as per the recommended preventive maintenance program, listed with part numbers and descriptions.
- On-site training for installation, use and maintenance is described as preferable, together with remote support.
- Easy to install: preassembled and skid-mounted, or containerized.
Where it applies
The generator section lists levels of use: sub-county hospital (Level 4), county referral (Level 5) and national referral hospital (Level 6). The specification sets requirements for a buyer's tender; it does not say who inspects or approves a plant. For the regulator's device list, see which Kenyan body regulates medical devices.
More questions
What purity does the Kenyan specification require?
The generator section says the oxygen shall be medical grade with purity of 93±3%, as defined in the European Pharmacopoeia monograph Oxygen, 93 per cent.
Does it say anything about hot climates?
The compressor section says it is suitable for use in ambient temperatures up to +45°C, and the generator section asks for continuous operation at 5 to 45 °C or more.
Sources
Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.
Related
- AnswersWhich Kenyan body regulates medical devices?
- AnswersDoes a PSA oxygen plant need oil-free compressors?
- AnswersDoes a PSA plant need a backup generator or UPS?
- AnswersHow is a PSA oxygen plant protected from dust?
- AnswersWhat should a PSA oxygen plant tender specify?
- AnswersHow do you write a brand-neutral PSA plant specification?