Oxygen and PSA

What pressure does a PSA oxygen plant deliver?

At what pressure does a PSA oxygen plant deliver oxygen to the hospital pipeline?

Short answer

The 2020 WHO interim specification for PSA plants (superseded in 2024), from its COVID-19 series, asked for 300 to 600 kPa (3 to 6 bar) continuous output; ULTRAOX SC and ST plants are listed at 6 bar, and NHS England's HTM 02-01 (2026) gives 400 kPa as the nominal pressure of oxygen pipelines.

SourcePressureWhat it refers to
WHO interim PSA specification (8 June 2020, COVID-19 series)300 to 600 kPa (3 to 6 bar); alarm below 3 barContinuous output pressure of the plant
Global Fund / BHI briefing note (4 March 2025)3.5 to 5 bar; about 4 barPSA oxygen at the end of the system, deliverable to wall outlets; pressure at the generator outlet, often suitable for delivery
PATH and CHAI (July 2021)50 to 55 psig (NFPA 99) or 400 to 500 kPa (ISO, piping only)Continuous output pressure in basic specifications
HTM 02-01 (2026), NHS England400 kPaNominal pressure of oxygen pipelines
AIGA 113/208 to 9 bar(g)Nominal pressure of a double-stage pipeline system, where a booster may be needed
ULTRAOX SC and ST datasheets6 bar oxygen outlet; 7.5 bar air inletEvery model in the tables

The Global Fund note says a PSA air compressor pressurizes the incoming air to 4.5 to 7 bar, and the 2020 WHO interim specification asked for a feed air compressor of at least 750 kPa. The ULTRAOX datasheet (2022) states that every 1 bar increase in compressor pressure adds 7% to total electricity consumption, and that the plant is designed to keep a continuous stable pressure in the oxygen network.

AIGA 113/20, adopted from EIGA Doc 195, requires a pressure reducing station whenever the produced or stored pressure is higher than the pipeline distribution pressure. It says a booster compressor could be needed to reach the nominal pipeline pressure of EN ISO 7396-1, usually only for a double-stage distribution system, where the nominal pressure is 8 to 9 bar(g). The Global Fund lists as a PSA benefit that oxygen leaves the generator at about 4 bar, often suitable for delivery, without an additional booster; where national standards require piping above 6 bar, it says a medium pressure booster can raise 4 to 6 bar up to 11 to 12 bar.

The 2020 WHO interim specification placed a gauge or sensor between the source and the line pressure control to monitor the output pressure. For pressures across the hospital network, see what pressures do medical gas pipelines run at.

More questions

Does a PSA plant need a booster to feed the pipeline?

Not usually. The Global Fund (2025) says PSA oxygen leaves the generator at about 4 bar, often suitable for delivery; AIGA 113/20 says a booster is usually only needed for double-stage pipelines at 8 to 9 bar(g).

Why not run the compressor at a higher pressure?

The ULTRAOX datasheet (2022) states that every 1 bar increase in compressor pressure adds 7% to total electricity consumption.

Sources

Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.

Reviewed: · Ultra Controlo engineering · Markdown version