Oxygen and PSA
Where should the air intake of a PSA oxygen plant be located?
Where should the air intake of a PSA oxygen plant be located to avoid exhaust fumes, dust and other contaminants?
Short answer
HTM 02-01 (2026), the NHS England guidance, says PSA compressors should meet its medical air rules, and AIGA 113/20 says the compressor suction shall never be near engine exhaust, ventilation, vacuum or scavenging discharges, and requires the plant's own waste gas vents to be directed outside, away from personnel and equipment.
HTM 02-01 Part A (2026), paragraph 6.117, says compressors for PSA systems should comply with the medical air requirements of its Chapter 7. Those ask for the compressor air intake to be located to minimise contamination from internal combustion engine exhausts and the discharge from vacuum systems, AGSS and ventilation systems, ducted where necessary, and say a minimum height of 5 m above ground level should ensure a reasonable quality of intake air. The full medical air rules are in where a medical air intake should be.
What to keep away from the intake
- Exhausts and smoke: Global Fund site considerations (14 May 2025) say sources of smoke and exhaust, including burn pits, incinerators, diesel generators and vehicle exhausts, must be more than 10 meters from the PSA plant. PATH (November 2023) lists a generator exhaust near the intake among the inappropriate installation methods it observed.
- The wider site: AIGA 113/20, adopted from EIGA Doc 195, says air quality is probably the major concern in selecting a suitable site, asks for site surveys over a significant area around the proposed location and attention to its future development, and, if the investigation is not definitive, for monitoring of the air at the site over an extended period. It recommends consulting the manufacturer on the acceptable levels of air contamination considered in the equipment design.
- Rain, insects and blockage: UNDP's site evaluation checklist (REQ-OXY-001, Annex II, 2025) asks for the distance from the air intake to the nearest contaminant source, and whether the intake is protected from insect entry and precipitation and free from blockage such as debris.
The plant's own exhaust
HTM 02-01 Part A, paragraph 6.110, says the waste product of a PSA plant, essentially nitrogen, is discharged to atmosphere during regeneration of the adsorbents. AIGA 113/20 notes that this waste gas, oxygen-depleted or oxygen-enriched air, can create oxygen-deficient or oxygen-enriched atmospheres around the vents and can carry dust from the adsorber beds. It requires all vents to be directed to a safe location outside the buildings, away from personnel and equipment, prohibits discharge at the inlet of an air conditioner, and asks for vents that are unrestricted and protected from damage by nearby activities such as vehicle traffic. UNDP (2025) asks for the nitrogen discharge duct to be vented at least 3 meters from doors or windows and to face downward.
For filters and the housing against dust, see how a PSA plant is protected from dust.
More questions
Do PSA compressors follow the medical air intake rules?
Yes, in NHS England's HTM 02-01 (2026): paragraph 6.117 says compressors for PSA systems should comply with the medical air requirements of Chapter 7.
How far should exhausts be from a PSA plant?
Global Fund site considerations (2025) say sources of smoke and exhaust, such as diesel generators and vehicle exhausts, must be more than 10 meters from the PSA plant.
Sources
- NHS England, HTM 02-01 Medical gas pipeline systems, Part A (2026), paragraphs 6.110, 6.117, 7.23 and 7.24
- Global Fund / Build Health International, Overview: Oxygen Plant Site Considerations (14 May 2025)
- PATH, Pressure swing adsorption plants: common issues and solutions to prevent future disrepair (November 2023)
- AIGA 113/20 (adopted from EIGA Doc 195), Safe design and operation of on site generation of oxygen 93% for medicinal use, sections 6.2.1 and 6.9
- UNDP, REQ-OXY-001 v01, Technical and regulatory requirements for oxygen plants (effective 1 January 2025), section 6.3 and Annex II
Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.