Standards
ISO 7396-1 vs HTM 02-01 vs NFPA 99: how do they differ?
Short answer
ISO 7396-1 is the international standard for medical gas and vacuum pipelines; HTM 02-01 is NHS England guidance with a design part and an operational management part; NFPA 99 is the US Health Care Facilities Code, a broader risk-based code. A 2023 WHO document names all three as reference standards widely used for pipeline design.
The three documents deal with the same engineering problem but come from different bodies, carry different legal weight and cover different ground. WHO's Foundations of medical oxygen systems (17 February 2023) names them as the three reference standards widely used for MGPS design.
| ISO 7396-1 | HTM 02-01 | NFPA 99 | |
|---|---|---|---|
| Publisher | ISO (ISO/TC 121/SC 6) | NHS England | National Fire Protection Association (NFPA) |
| Type of document | International standard; adopted in Europe as EN ISO 7396-1:2016 + A1:2019 | Health Technical Memorandum, published for England; not mandatory unless specifically stated | Health Care Facilities Code |
| Edition on 22 October 2026 | ISO 7396-1:2016 + Amd 1:2017; a 4th edition is under publication | 2026 edition, published 19 August 2026, in Part A and Part B | 2027 edition |
| Scope | Design, installation, function, performance, testing, commissioning and documentation of pipeline systems for medical gases and vacuum | Design, installation, validation and verification (Part A) and operational management of the MGPS and cylinders (Part B) | Criteria for health care services and systems based on risk to patients, staff and visitors, to minimise fire, explosion and electrical hazards |
ISO 7396-1: the international base
ISO 7396-1 sets requirements for supply systems, pipeline distribution, control, monitoring and alarm systems, and non-interchangeability between gas and vacuum services. Oxygen 93 is in scope: where every source delivers oxygen 93, the concentration may vary between 90% and 96%. Anaesthetic gas scavenging disposal systems are left to ISO 7396-2, and the 2016 edition does not apply to dental vacuum or to cylinder filling systems.
HTM 02-01: guidance with an operational part
HTM 02-01 follows BS EN ISO 7396-1 on supply: every medical gas supply should have three sources, called primary, secondary and tertiary. Part B, on operational management, names roles such as the Authorised Person (MGPS), Competent Person (MGPS) and Quality Controller (MGPS). It treats medical gases as medicines under the Human Medicines Regulations 2012 and the British Pharmacopoeia, treats medical air as a 400 kPa service and asks for surgical air of at least 700 kPa at the terminal unit, and says PSA systems should only be considered where geographical or environmental constraints prevent liquid oxygen or compressed gas supply.
NFPA 99: a whole-facility code
NFPA 99 is wider than a pipeline standard. NFPA describes it as setting criteria for levels of health care services or systems based on risk, and its technical committees include piping systems, electrical systems, mechanical systems, medical equipment, hyperbaric and hypobaric facilities, and emergency management and security. NFPA lists the 2027 edition as current; older editions still appear in specifications, such as the 2018 edition referenced in the WHO PSA plant specification of 2020 (superseded by WHO's technical specifications of 31 December 2024). For people, IAPMO states that the ASSE/IAPMO/ANSI Series 6000-2024 qualification standard for medical gas systems personnel aligns with the medical gas requirements of NFPA 99-2024.
Which one applies to a project
WHO's Foundations of medical oxygen systems (2023) notes that national authorities indicate the applicable local standard. In Europe ISO 7396-1 is adopted as EN ISO 7396-1:2016 + A1:2019, HTM 02-01 is published for England, NFPA 99 is a US code, and in Canada CSA Group describes CSA Z7396.1:22 as the governing standard for medical gas installations in health care institutions. Whichever is used, check the edition in force: HTM 02-01 of 2006 and HTM 2022 have been superseded, and a 4th edition of ISO 7396-1 is under publication.
The ULTRAOX SC datasheet states compliance with ISO 7396-1:2016, and the ULTRAAR SC datasheet states that the medical air plant is built to EN ISO 7396-1 and fulfils HTM 02.
More questions
Is HTM 02-01 mandatory?
The HTM states that its guidance is not mandatory unless specifically stated, but any departure should provide a level of safety not less than that achieved by following it.
Which edition of NFPA 99 is current?
NFPA lists the 2027 edition as current. The WHO PSA plant specification of 2020 still referred to the 2018 edition.
Does ISO 7396-1 cover anaesthetic gas scavenging?
No. Anaesthetic gas scavenging disposal systems are covered by ISO 7396-2.
Sources
- ISO 7396-1:2016, iso.org
- ISO 7396-1, 4th edition (under publication), iso.org
- DIN EN ISO 7396-1:2019-06 (EN ISO 7396-1:2016 + A1:2019), DIN Media
- NHS England, Health Technical Memorandum 02-01 (publication page)
- HTM 02-01 Part A: Design, installation, validation and verification (2026)
- HTM 02-01 Part B: Operational management (2026)
- NFPA 99, Health Care Facilities Code (document information page)
- WHO, Foundations of medical oxygen systems (17 February 2023)
- WHO, Technical specifications for PSA oxygen plants: interim guidance (8 June 2020)
- IAPMO, ASSE/IAPMO/ANSI Series 6000-2024 now available (press release)
- CSA Group, CSA Z7396.1:22
- ULTRAOX SC datasheet, p. 2
- ULTRAAR SC datasheet, p. 2
- WHO, Technical specifications for health facility based medical oxygen systems (31 December 2024), note on the oxygen generator plant specification (supersedes the 2020 interim PSA specifications)
Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.
Related
- StandardsISO 7396-1: pipelines for compressed medical gases and vacuum
- StandardsHTM 02-01: medical gas pipeline systems (NHS England)
- StandardsNFPA 99: Health Care Facilities Code for medical gases
- AnswersWhat changed in the 2026 edition of HTM 02-01?
- AnswersWhat is a medical gas pipeline system (MGPS)?