How often should a PSA plant's oxygen analyser be checked?

How often should the oxygen analyser of a PSA plant be checked against a handheld analyser?

Short answer

Global Fund guidance asks for a daily check: its PSA daily checklist records the purity on the plant's PLC alongside a handheld analyser reading, and its 2024 sustainability guidance asks for purity to be recorded every day with at least two different analysers.

Check frequencies in the guidance

WhenCheckSource
Each day the plant runsRecord the purity shown on the PLC and the purity measured with a handheld analyserGlobal Fund daily maintenance checklist
DailyRecord purity with at least two different analysers; at a minimum, measure it with a handheld analyserGlobal Fund sustainability guidance (2024)
DailyInspect oxygen purity, with the pre-filter, compressor pressure and oil level, tank pressures and dryer dew pointUSAID RISE operating procedure, written for hospitals in India
At least quarterlyTest the quality of gas from the PSA plantHTM 02-01 Part B, NHS England (2026)
Preferably every six monthsHave oxygen samples tested by an accredited laboratory to validate the plant and its oxygen analyserUSAID RISE operating procedure
YearlyCalibrate the oxygen sensorUSAID RISE operating procedure
At commissioningVerify the operation of the oxygen analyser and all alarmsWHO interim technical specifications for PSA plants (2020, superseded by WHO's technical specifications of 31 December 2024)

When the readings disagree

The USAID RISE procedure lists a faulty sensor, and a faulty PLC reading that must be calibrated as the manufacturer recommends, among the points to check when purity is low. HTM 02-01 asks for connections for calibration cylinders at the plant, alongside its duplex monitoring system, and PATH and CHAI (2021) include calibrating pressure and oxygen sensors in the preventive maintenance a buyer can contract to the manufacturer. The 2020 WHO interim specification asked for the plant's oxygen concentration monitor to have ±1% accuracy.

As reference points, Global Fund's 2025 guide to expected values treats above 90% as medical grade and says that a reading below 85% should lead to an investigation, while HTM 02-01 isolates a PSA plant from the pipeline below 94%. Open Critical Care says periodic checks need oxygen analysers and suggests asking the manufacturer or authorised distributor for the specific analysers needed for maintenance, and WHO (2023) lists an oxygen analyser in the maintenance toolkit that operators should have.

Records

Global Fund's checklist asks the technician who completes it to initial the form, and asks for a separate set of records for each system in a duplex or triplex plant.

Ultra Controlo analysers

Ultra Controlo supplies a portable oxygen purity analyser for on-site checks of oxygen concentration in pipelines and cylinders. On ULTRAOX SC plants, purity is measured by one or two oxygen analysers with a zirconium oxide probe, and the QuPHARM wall-mounted analyser, launched in May 2026, has programmable alarm thresholds on its oxygen, carbon monoxide and carbon dioxide channels.

More questions

Which reading should prompt action?

Global Fund's expected values (2025) say a reading below 85% should lead to an investigation; NHS England's HTM 02-01 isolates the PSA plant from the pipeline below 94%.

Who keeps the purity records?

Global Fund's checklist asks the technician who completes it to initial each day's form, and asks for a separate set of records for each system of a duplex or triplex plant.

Sources