Oxygen and PSA

What did a Chilean study find about a PSA oxygen plant at 5,050 m?

What did a 2021 study of the ALMA observatory in Chile report about a PSA oxygen plant installed at 5,050 m?

Short answer

A 2021 Frontiers in Physiology study by staff of the ALMA observatory and Chilean universities reports that a PSA plant installed in 2008 at the 5,050 m Chajnantor site, rated at 850 l/min of 90 to 95% oxygen, suffered oil contamination and operating problems, and that oxygen enrichment from it was stopped in 2011.

This is a field report from an observatory, not a hospital. The plant enriched the air of work buildings, not patient rooms, so it describes one PSA installation at very high altitude, not how a hospital plant performs.

The installation

The authors say the Atacama Large Millimeter/submillimeter Array operates at 5,050 m on the Chajnantor Plateau in Chile. In 2008 it installed an onsite PSA oxygen generator to enrich the air of all rooms in its Array Operations Site buildings to 28% oxygen, which the paper equates to an altitude of 2,900 m. The system has two external tanks of 2,498 litres each and a maximum oxygen pressure of 75 psi, delivering 850 l/min with 90 to 95% oxygen purity at 45 psi outlet pressure.

What went wrong

The paper reports operational troubles after installation, including not reaching the required oxygenation. An inspection after shutdown found a series of malfunctions, starting with oil contamination along the air preparation route, which the authors call unacceptable because of the risk of rapid self-ignition. It also says the zeolite beds, the oxygen storage tanks and the pipelines needed to be exchanged because of oil and water contamination. In 2011 ALMA stopped the oxygen administration in the building areas because of problems in the operation of the PSA plant. In 2015 a liquid oxygen tank was installed.

What the authors say about altitude

The authors cite an earlier paper (Litch and Bishop, 2000) for two points: regular maintenance is needed, including changing filters and zeolite columns after 15,000 to 20,000 hours of use depending on local conditions, and output oxygen concentration at high altitude falls by about 10% for each 2,000 m of elevation gained. From that rule of thumb they say they expect a 25% decrease at 5,050 m. This is their expectation, not a measurement from the ALMA plant, and the rule it rests on comes from another paper. They conclude that the PSA plant needs adequacy and maintenance to operate at very high altitude.

The study does not give a figure for hospitals. For how altitude is handled in specifications, see how altitude affects a PSA plant.

More questions

Was this plant in a hospital?

No. It served the work buildings of the ALMA observatory at 5,050 m, where the target was 28% oxygen in room air.

Does the paper measure how much the plant lost at altitude?

No. The 25% decrease is the authors' expectation, based on a rule of 10% per 2,000 m that they cite from an earlier paper.

Sources