Medical gas library

Oxygen and PSA

Floor, doors and finishes of a PSA plant room

What floor, doors, lighting and finishes should the room that houses a PSA oxygen plant have?

Short answer

WHO's Foundations of medical oxygen systems (2023) says the doors of a PSA plant room should have external protective shutters that only authorised personnel can access, with signage for medicinal gases and fire safety, and that the floor should be flat, smooth, easy to clean and strong enough for the equipment.

WHO's Foundations of medical oxygen systems (17 February 2023, sections 2.2.4 and 3.5.1) and UNDP's requirements for oxygen plants (REQ-OXY-001, effective 1 January 2025, section 6.3) give a similar list for the plant room. WHO states that its structural considerations are guidance only and that further consultation with civil engineers, architects and other experts must be performed.

Doors and access

  • WHO: the entrance doors should have external protective shutters, which only authorised personnel can access, and appropriate signage for medicinal gases and fire safety. UNDP says the same in its own words.
  • WHO's list of architectural elements also includes doors wide enough for the passage of equipment.

Floor and finishes

  • Floor: WHO says the finish and material, for example concrete, must be flat and smooth to ease movement and cleaning, sustain the heavy weight of the components and reduce equipment vibration. UNDP asks for floors that are flat, durable, easy to clean and vibration-reducing.
  • Slabs: WHO lists concrete slabs designed for the load of the equipment, given in kg/m2, with a broom finish and a non-slip finish.
  • Finishes: easy to clean and kept dust free, and WHO suggests grids or nets in openings such as windows to keep the room as clean as possible.
  • Below ground: WHO notes that underground rooms, pits, vessels, ducts and trenches may represent a risk for oxygen accumulation.

Drainage and lighting

  • Ducts for the water discharged by the water separator.
  • Appropriate continuous lighting in the plant room: WHO notes that LED lighting reduces fire risks. It also asks for proper exterior and interior lighting during the whole day.

Ventilation and the nitrogen vent are covered in what ventilation a PSA plant room needs; size and layout in how much space a PSA plant needs; weather protection and roofs in whether a PSA plant can be installed outdoors or on a roof.

More questions

Does WHO give a floor load figure for the plant room?

Not in the text that lists the plant room requirements: WHO says concrete slabs should be designed considering the load of the equipment, in kg/m2. UNDP's site evaluation checklist asks whether the slab load resistance is calculated for a minimum of 1000 kg/m².

Sources

Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.

Reviewed: · Ultra Controlo engineering · Markdown version