Medical gas library

Pipeline systems

Double-stage pipeline distribution

In AIGA 113/20, a double-stage pipeline distribution system is one in which gas is initially distributed from the supply system at a pressure higher than the nominal distribution pressure and then reduced to it by line pressure regulators.

AIGA 113/20, adopted from EIGA Doc 195, says a booster compressor could be required to ensure that the supply gas is at the nominal pressure of the pipeline according to EN ISO 7396-1, usually only where a double-stage system is used, with a nominal pressure of 8 to 9 bar(g).

Under HTM 02-01 (2026), the NHS England guidance, medical gases may be distributed at a higher pressure than the nominal pressure at terminal units. The maximum working pressure should then not exceed 1100 kPa and a pressure-reducing station (PRS) should be installed. Simplex systems should not be used, and the PRS should include duplex pressure-regulating valves.

Portugal's ACSS ET 03/2006 (2024 revision) gives a primary network pressure of 800 kPa ± 100 for oxygen and 480 kPa + 100 in the secondary oxygen network, and says each service should be supplied by at least two second-reduction sets in bypass. These are Portuguese values, not a general rule.

The ULTRALOBO LPT datasheet says Ultra Controlo developed its low-pressure oxygen compression systems to meet the needs of hospitals without altering existing medical oxygen networks or compromising approved local standards. See how a hospital can use PSA oxygen without changing its network.

Sources

Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.