Oxygen and PSA
Does a PSA oxygen plant need a booster compressor?
Does a PSA oxygen plant need a booster compressor, and in which cases is one added?
Short answer
The Global Fund briefing note on PSA, VPSA and VSA (4 March 2025) says PSA oxygen leaves the generator at about 4 bar, often suitable for delivery, so no additional booster is needed; a booster is added to fill cylinders, for VPSA and VSA plants, or where national standards require piping above 6 bar.
In the Global Fund briefing note, a PSA air compressor pressurizes the incoming air to 4.5 to 7 bar, and the oxygen pressure at the end of the system is high enough to be delivered directly to wall-mounted oxygen outlets (3.5 to 5 bar). The note lists as a PSA benefit that oxygen exits the generator at about 4 bar, often suitable for delivery pressure, so an additional oxygen booster is not needed. For the pressures that other documents give, see what pressure a PSA plant delivers.
When the sources say a booster is added
- To fill cylinders. The Global Fund note says PSA oxygen can be fed directly to a high pressure booster to fill cylinders, and WHO's Foundations of medical oxygen systems (2023) describes that booster as raising the gas from 3 to 6 bar to up to 200 bar. See what a high-pressure oxygen booster is.
- On VPSA and VSA plants. The Global Fund note says the oxygen pressure at the end of a VSA system is not high enough for wall outlets, and lists an additional oil-free oxygen booster compressor for distribution among the drawbacks of VPSA and VSA. WHO (2023) says the VSA output pressure, from 130 to 420 kPa, is below the common requirements for pipeline networks, which generally adds the need for an oil-free booster compressor to attain 689 kPa. PATH and CHAI (2021) say VSA plants often require an additional high-pressure oxygen compressor for piping networks, and UNDP's REQ-OXY-001 (2025), describing VSA, says a booster compressor is required for output pressure. See can VSA oxygen be used in hospitals.
- Where national standards require piping above 6 bar. The Global Fund note says an additional medium pressure booster can then raise the oxygen from 4 to 6 bar up to 11 to 12 bar. AIGA 113/20, adopted from EIGA Doc 195, says a booster compressor could be required to reach the nominal pipeline pressure of EN ISO 7396-1, usually only where a double-stage pipeline distribution system is used, with a nominal pressure of 8 to 9 bar(g). See double-stage pipeline distribution.
Ultra Controlo boosters
The ULTRALOBO LPT datasheet says Ultra Controlo developed its low-pressure oxygen compression systems to meet the needs of hospitals without altering existing medical oxygen networks or compromising approved local standards, and its model table lists a 5 bar inlet and a 10 to 12 bar maximum outlet for every model. See the ULTRALOBO LPT range and how it uses PSA oxygen in an existing network. For cylinders, the ULTRACECO HP datasheet describes compressors suitable for compressing oxygen in 170/200 bar cylinders. The two jobs are compared in low-pressure vs high-pressure oxygen boosters.
More questions
Does a PSA plant that fills cylinders need a booster?
The Global Fund note says PSA oxygen can be fed directly to a high pressure booster to fill cylinders, and WHO (2023) defines a cylinder filling station as a booster compressor and a cylinder filling ramp.
Does the booster have to be oil-free?
WHO (2023) says the booster compressor used to fill cylinders must be an oil-free compressor, and the Global Fund note lists an oil-free oxygen booster compressor for VPSA and VSA distribution.
Sources
- Global Fund / Build Health International, Briefing Note: PSA, VPSA and VSA (4 March 2025)
- WHO, Foundations of medical oxygen systems (17 February 2023), glossary, 2.1 and 2.5
- AIGA 113/20 (adopted from EIGA Doc 195), On site generation of oxygen 93% for medicinal use, 4.3.1
- UNDP, REQ-OXY-001 v01 (2025), 6.1
- PATH and CHAI, Oxygen Generation and Storage: PSA/VSA (2021)
- ULTRALOBO LPT datasheet, pp. 2 and 4
- ULTRACECO HP datasheet, p. 2
Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.
Related
- AnswersWhat pressure does a PSA oxygen plant deliver?
- AnswersCan VSA oxygen plants be used in hospitals?
- ProductsULTRALOBO LPT: specifications and models
- AnswersCan a containerised oxygen plant be moved?
- AnswersCan a hospital sell surplus oxygen to nearby facilities?
- AnswersCan a hospital with broken gas pipelines use a PSA plant?