Oxygen and PSA
Low-pressure vs high-pressure oxygen boosters
Short answer
The Global Fund briefing note (4 March 2025) describes a medium pressure booster that raises oxygen from 4 to 6 bar up to 11 to 12 bar for piping, and WHO (2023) a high-pressure booster that raises it from 3 to 6 bar to up to 200 bar to fill cylinders.
The sources use different names for boosters that do different jobs. The Global Fund (2025) says a medium pressure booster can be added where national standards require piping above 6 bar, and it also speaks of a high pressure booster that fills cylinders. WHO (2023) calls the cylinder-filling device a booster compressor or high-pressure booster, and AIGA 113/20 calls it a high pressure compressor. The Ultra Controlo ULTRALOBO LPT datasheet describes its boosters as low-pressure.
Side by side
| Booster for the pipeline | Booster for cylinders | |
|---|---|---|
| Job described in the sources | Raises oxygen to the pipeline pressure: where national standards require piping above 6 bar (Global Fund, 2025), or for a double-stage pipeline system (AIGA 113/20) | Fills high-pressure cylinders: WHO (2023) defines a cylinder filling station as a booster compressor and a cylinder filling ramp |
| Pressure in the sources | 4 to 6 bar up to 11 to 12 bar (Global Fund, 2025); a nominal pipeline pressure of 8 to 9 bar(g) in a double-stage system (AIGA 113/20) | 3 to 6 bar up to 200 bar (WHO, 2023); re-filling reserve vessels up to 200 bar(g), depending on national regulations (AIGA 113/20) |
| Name the source uses | Medium pressure booster (Global Fund); booster compressor (AIGA 113/20) | High pressure booster (Global Fund); booster compressor or high-pressure booster (WHO); high pressure compressor (AIGA 113/20) |
| Oil | The Global Fund lists an additional oil-free oxygen booster compressor for VPSA and VSA distribution | WHO (2023): must be an oil-free compressor |
| Ultra Controlo datasheet | ULTRALOBO LPT: ULTRACECO LPT low-pressure boosters, a completely oil-free system, all parts in contact with oxygen in AISI 316 stainless steel, and a model table with a 5 bar inlet and a 10 to 12 bar maximum outlet | ULTRACECO HP: oil-free, suitable for compressing oxygen in 170/200 bar cylinders. ULTRACECO HPT: without oil, optionally up to 3,000 PSI (207 bar g). ULTRACECO HPX: oil-free |
The figures come from different documents and describe different equipment, so they are not a like-for-like comparison. For the ULTRALOBO LPT, the datasheet table gives the inlet and maximum outlet of each model; see the ULTRALOBO LPT range. For when a PSA plant needs either kind, see does a PSA plant need a booster; for the cylinder side, see ULTRACECO HP, HPT and HPX compared.
More questions
Is a medium pressure booster the same as a low-pressure booster?
The documents use different names: the Global Fund note says medium pressure booster and the ULTRALOBO LPT datasheet says low-pressure boosters. Each name is used here with the document that uses it.
Does a booster that fills cylinders have to be oil-free?
WHO (2023) says the booster compressor, also known as a high-pressure compressor, must be an oil-free compressor.
Sources
- Global Fund / Build Health International, Briefing Note: PSA, VPSA and VSA (4 March 2025)
- WHO, Foundations of medical oxygen systems (17 February 2023), glossary and 2.5
- AIGA 113/20 (adopted from EIGA Doc 195), On site generation of oxygen 93% for medicinal use, 4.3.1 and 4.3.2
- ULTRALOBO LPT datasheet, pp. 3 and 4
- ULTRACECO HP datasheet, p. 2; ULTRACECO HPX datasheet, p. 2 (no separate product page)
- ULTRACECO HPT datasheet, p. 2
Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.
Related
- ProductsULTRALOBO LPT: specifications and models
- AnswersDoes a PSA oxygen plant need a booster compressor?
- ComparisonsAir-cooled vs water-cooled oxygen booster
- ComparisonsContainerised vs plant-room oxygen plant
- ComparisonsGalvanic vs ultrasonic handheld oxygen analysers
- ComparisonsOn-site oxygen vs delivered oxygen: how do costs compare?