Oxygen and PSA
How long does an oxygen analyser sensor last?
How long does the galvanic cell sensor of a handheld oxygen analyser last before it must be replaced?
Short answer
WHO-UNICEF's guidance on oxygen therapy devices (2019) says a galvanic cell oxygen sensor should typically last at least 12 months, while its specification asks for a nominal operating life of at least 1.5 years or 600 000 oxygen percentage hours, whichever is greater.
These figures come from chapter 8 of the WHO-UNICEF Technical specifications and guidance for oxygen therapy devices (2019), whose focus is portable handheld analysers for spot checks. It is a specification for oxygen therapy devices, not a guide to the fixed analysers of a PSA plant. The chapter gives several figures in different places and does not reconcile them, so they are listed here as written.
| Where in the chapter | What it says |
|---|---|
| Overview (8.1) | A galvanic sensor should typically last at least 12 months. |
| Table 8.1, drawbacks of the electrochemical (galvanic cell) analyser | Lifetime: less than 1 to 2 years. |
| Specification (8.2.1) | Replaceable galvanic cell with a nominal operating life of at least 1.5 years or 600 000 oxygen percentage hours, whichever is greater. |
| Maintenance (8.2.5) | Replace the sensor as per the manufacturer's recommendation, usually every 6 to 12 months. |
Why the sensor wears out
WHO-UNICEF says that, because of the accumulation of electrolyte residue and oxidation over time, galvanic sensors need to be periodically replaced. An inability to calibrate the analyser usually means that the sensor needs to be changed. After replacing the sensor, WHO-UNICEF says to calibrate the analyser with a known oxygen concentration level. See how a handheld oxygen analyser should be used and cared for.
Where handheld analysers sit in WHO's plant guidance
WHO's Foundations of medical oxygen systems (2023) says a maintenance toolkit with an oxygen analyser, other testing and mechanical tools and complementary PPE should be available for the operators of a plant, and WHO's 2024 specification for an oxygen generator plant lists a hand-held (stand-alone) oxygen analyser for quality control of product among the other system requirements to consider in procurement.
Ultrasonic analysers
Table 8.1 lists, as merits of the ultrasonic analyser, no need to replace the sensor and a lifetime of at least 5 years. See galvanic and ultrasonic oxygen analysers compared.
More questions
What does it mean if a handheld oxygen analyser cannot be calibrated?
WHO-UNICEF (2019) says an inability to calibrate the analyser usually means that the sensor needs to be changed.
Do ultrasonic oxygen analysers need new sensors?
Table 8.1 of WHO-UNICEF (2019) lists no need to replace the sensor and a lifetime of at least 5 years as merits of the ultrasonic analyser.
Sources
- WHO-UNICEF, Technical specifications and guidance for oxygen therapy devices (2019), chapter 8, sections 8.1, 8.2.1 and 8.2.5 and Table 8.1
- WHO, Foundations of medical oxygen systems (17 February 2023), section on plant maintenance
- WHO, Technical specifications for health facility based medical oxygen systems (31 December 2024), oxygen generator plant table, section 1.1.1
Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.