How should a handheld oxygen analyser be used and cared for?

How should a handheld galvanic cell oxygen analyser be used, calibrated, stored and cleaned?

Short answer

WHO-UNICEF's guidance on oxygen therapy devices (2019) says not to connect a handheld oxygen analyser to the high-pressure side of an oxygen cylinder, to run a functional test before using the device, and to calibrate it, including after a sensor replacement, at the pressure of the gas to be measured.

This page follows chapter 8 of the 2019 WHO-UNICEF Technical specifications and guidance for oxygen therapy devices, whose focus is portable handheld analysers used to test oxygen from a concentrator or the low-pressure side of a compressed gas cylinder or terminal unit source. The safety, handling and maintenance points below come from its section 8.2, which is written for the galvanic cell analyser.

Before and while measuring

  • Pressure: the specification asks for an analyser suitable for measuring gas supply with pressure up to 345 kPa (3.45 bar, 50 psi), and WHO-UNICEF says not to connect it to the high-pressure side of an oxygen cylinder.
  • Functional test: users should run one before using the device. WHO-UNICEF also says to follow the manufacturer's instruction manual on installation.
  • Ultrasonic analysers: humidifiers should be removed before testing oxygen concentration from an oxygen source, because high humidity can damage the sensor.

Calibration

WHO-UNICEF says galvanic cell calibration is usually accomplished by exposing the electrode to room air (21% oxygen) and another source of known concentration, for example 100% oxygen from a cylinder. Because calibration depends on the gas pressure, it must be carried out at the pressure of the gas that will be measured. After a sensor replacement, it gives common practice as 21% from the air and/or 100% from a cylinder filled by a cryogenic plant.

Handling the sensor

  • Wear rubber gloves while replacing the galvanic cell sensor.
  • Do not remove the cell and expose it to air unless you have an indication that it is not performing.
  • The cell has thin sensing membranes: do not puncture them with sharp objects. The sensor must be replaced if the membrane is non-functioning.

Where handheld analysers sit in WHO's plant guidance

WHO (2023) says a maintenance toolkit with an oxygen analyser, other testing and mechanical tools and complementary PPE should be available for the operators of a plant, and WHO's 2024 specification for an oxygen generator plant lists a hand-held (stand-alone) oxygen analyser for quality control of product among the other system requirements to consider in procurement.

Cleaning, storage and faults

WHO-UNICEF says to clean the exterior with a cloth very lightly dampened with a mild detergent and water, then wipe dry with a clean cloth. When not in use, the device should be stored in a secure and clean location, in its storage case, with controlled temperature and humidity and protected from dust and insects. For faults, follow the manufacturer's service manual, which shows the error code, probable cause and proposed remedial action; otherwise contact the manufacturer's first-line support team. For the sensor replacement interval, see how long an oxygen analyser sensor lasts.

More questions

Can a handheld oxygen analyser be connected to a cylinder valve?

WHO-UNICEF (2019) says not to connect the analyser to the high-pressure side of an oxygen cylinder; its focus is the low-pressure side of a cylinder or terminal unit source.

Why must a galvanic analyser be calibrated at the gas pressure?

WHO-UNICEF (2019) says calibration is dependent upon the gas pressure, so it must be carried out at the pressure of the gas that will be measured.