How to size a vacuum pump for pump-down or continuous flow

How do you size a vacuum pump for a pump-down time or a continuous flow?

Short answer

ISO 21360-1:2020 specifies three methods for measuring a vacuum pump's volume flow rate: the throughput method, a steady gas flow injected at a measured inlet pressure; the orifice method; and the pump-down method, calculated from two pressures, the pumping interval and the vessel volume. Leaks, desorption and line resistance change the result.

Pump-down duty

In the pump-down method of ISO 21360-1:2020, last confirmed by ISO in 2025, the volume flow rate is calculated from the pressures before and after a pumping interval and from the volume of the test vessel. A pump-down duty is described by the same data, so a request to a supplier should give the volume to be evacuated, including pipework, the starting and target pressures in absolute terms, and the time allowed. The standard names leak and desorption rates, gas cooling during the pumping interval and increasing flow resistance in the connection line at low pressures as effects that influence the measured volume flow rate.

Continuous flow duty

In the throughput method of the same standard, a steady gas flow is injected into the pump while the inlet pressure is measured. For a continuous duty, the matching data are the gas flow that the process releases or that leaks in, and the absolute pressure at which the system is to be held. Elmo Rietschle, for example, designs its SVT screw pumps to deliver stable vacuum at 100 to 250 mbar(abs) for central vacuum, and says the SVT controller coordinates up to 8 pumps with trim and base-load logic.

Reading the maker's curves

  • Inlet conditions: Thomas states that its flow values relate to air at the pump inlet condition, that is, expanded under vacuum.
  • Reference conditions and tolerance: Elmo Rietschle gives curves and tables with a tolerance of ±10 % according to PNEUROP standards, for a pump at normal operating temperature (1 bar(abs.) and 20 °C), in datasheet D281; Speck Pumpen (Circutec) gives -10% on suction capacity and +10% on required output for its liquid ring pumps (process industry brochure, 06/2018).
  • Supply frequency: nominal suction capacity is stated for a frequency, for example 50 Hz in Elmo Rietschle datasheet D236.
  • Vapour: Elmo Rietschle says its L-BV5 liquid ring pump can act as a condenser, doubling the suction volume when dealing with condensable vapour; ISO 21360-2:2020 adds water vapour tolerance to the quantities measured on positive displacement pumps.

Which technology to size is covered in how to choose an industrial vacuum pump and what vacuum level a process needs.

Ultra Controlo distributes and services Elmo Rietschle vacuum pumps and Thomas pumps in Portugal as partner brands, and its UltraCare programme covers preventive and corrective maintenance, original spare parts and emergency support for partner-brand equipment.

More questions

Which standard defines how vacuum pump flow is measured?

ISO 21360-1:2020 specifies three methods for measuring volume flow rate, the throughput, orifice and pump-down methods, and one method each for base pressure, compression ratio and critical backing pressure.

Can several vacuum pumps share a varying load?

Elmo Rietschle says its SVT controller coordinates up to 8 pumps with trim and base-load logic for plant-wide efficiency.