How much seal water does a liquid ring vacuum pump need?

How much operating (seal) water does a liquid ring vacuum pump need?

Short answer

The US EPA guide WaterSense at Work (October 2012) puts the sealing and cooling water of a once-through liquid ring vacuum pump at 0.5 to 1.0 gallons per minute per horsepower, and estimates that full recirculation cuts water use by 80 percent and partial recirculation by about 50 percent.

The guide is written for commercial and institutional facilities. It gives the 0.5 to 1.0 gpm per hp range as the water needed to create the vacuum and cool the equipment, and estimates that full recovery and recirculation cuts water use by 80 percent and partial recovery and recirculation by about 50 percent. It also says the pump should be set at manufacturer specifications, so that it discharges only the water needed to remove impurities and cool it.

Three ways to supply the liquid

US EPA, WaterSense at Work, section 7.3 (October 2012), pp. 7-10 to 7-13.
ArrangementWhat happens to the liquidWater use the guide gives
Once-throughSeal and cooling water are continuously discharged and replenished with fresh water0.5 to 1.0 gpm per hp, for creating the vacuum and cooling
Partial recovery and recirculationPart of the sealing water is recirculated; make-up water keeps the impurity concentration from getting too high, and heat buildup in the pump should be avoidedAbout 50 percent less
Full recovery and recirculationAll the seal water is recovered from the discharge side, passed through a heat exchanger where the configuration allows, and reused; a small amount is discharged to remove impurities and make-up water is addedAn estimated 80 percent less

What the makers add

  • Elmo Rietschle says the operating liquid has a central role in sealing, cooling and condensing, so liquid management is essential: separators, circulation, replenishment and treatment where required.
  • The expelled liquid can be recirculated through the maker's standardised circuit units, which the maker says reduces water usage and operating costs.
  • The compact L-BL2 unit has a closed water circuit with no need to supply fresh water, and the maker says its patented discharge-air cooling system limits the temperature rise to no more than 15 K above room temperature.
  • Temperature sets the vacuum: as the operating liquid warms, its vapour pressure rises and the achievable ultimate vacuum decreases, so Elmo Rietschle advises cooling the liquid or using a circuit with a heat exchanger.
  • Speck Pumpen (Circutec) rates its liquid ring pumps for operating liquid up to 80 °C (process industry brochure, 06/2018) and measures its characteristic curves with water at nominal speed as per DIN 28431.

Measuring an existing pump

The EPA guide estimates current use with Equation 7-1: water use in gallons per year equals the discharge flow rate of the pump times the daily use time times the days of operation. Its operating tips are to turn the pump off when it is not in use or needed, to set it at manufacturer specifications, and to check control schemes such as timers, float-operated switches and total dissolved solids controllers. A recovery and recirculation system should be maintained per manufacturer instructions so that impurities are removed and hard water deposits do not remain.

Dry alternatives

The guide says a non-lubricated, air-cooled dry vacuum pump can eliminate the pump's water use altogether, but that where explosive or corrosive gases are removed, a facility might only be able to consider a liquid ring pump. See claw, rotary vane or liquid ring and how a liquid ring pump works.

Ultra Controlo distributes and services Elmo Rietschle and Speck Pumpen (Circutec) in Portugal as partner brands, and its UltraCare programme covers preventive and corrective maintenance, original spare parts and emergency support for partner-brand equipment.

More questions

Does recirculating the seal water affect the vacuum?

The EPA guide says partial recirculation systems should avoid heat buildup in the pump, and Elmo Rietschle says the achievable ultimate vacuum decreases as the operating liquid temperature rises.

How is the water use of an existing pump estimated?

With Equation 7-1 of the EPA guide: discharge flow rate times daily use time times days of operation, in gallons per year.