How often should medical air dew point and CO sensors be calibrated?
How often should the dew point and carbon monoxide measuring equipment of a medical air system be calibrated?
Short answer
Under HTM 02-01 (2026), the NHS England guidance, plant dew point hygrometers are specified by accuracy (±3 °C, set to −46 °C), while the electronic analysers used for quality tests, including carbon monoxide, should be calibrated on site with certificated gases before testing and again once all tests are completed.
HTM 02-01 (2026) deals with two sets of instruments for medical air: the hygrometers that watch each dryer continuously, and the analysers the Quality Controller (MGPS) uses to test the delivered air, including for carbon monoxide.
Plant dew point hygrometers
Under the HTM, each dryer assembly should have its own dew point sensor and test point after the bacteria filter, so that the dryer can be tested before it is brought online and the product can be monitored continuously. The dryer control unit should contain hygrometers and displays with a minimum accuracy of ±3 °C over a range of at least −20 °C to −80 °C, set to −46 °C atmospheric dew point, and an automatic changeover to the standby dryer when the duty dryer fails on dryness or pressure. For compressor plant, Part B of the HTM says the testing and servicing procedures supplied by the manufacturer should be followed.
Quality control analysers, including carbon monoxide
Carbon monoxide is not in the HTM 02-01 list of plant faults; it is checked by quality control testing of terminal units, with a limit of 5 ppm (v/v). The HTM asks for electronic equipment for these tests because of its repeatability, resolution and accuracy, calibrated on site before testing with certificated calibration gases. If readings drift during testing, the analyser calibration should be checked and recalibrated and the Quality Controller should consider repeating a sample of tests; calibration should also be carried out once all tests are completed, and records should be kept of all calibrations. The Quality Controller (MGPS) should ensure that all equipment used for medical gas testing is maintained and serviced according to the manufacturer's recommendations and local protocols.
Contractor test equipment
Separately, HTM 02-01 Part B makes the maintenance contractor provide the test equipment for its tasks, identified in the method statement and in Chapter 19 of Part A. That equipment should be calibrated in accordance with the manufacturers' recommendations, and in any case against recognised national standards at a frequency of no less than annually, with calibration records kept by the contractor and produced for inspection by the Authorised Person (MGPS).
How often the air itself is tested
Quality control testing is carried out on new installations and after repairs and modifications, and the HTM sets quarterly testing for medical and surgical air compressor plants. Water vapour should not exceed 67 vpm, about −46 °C atmospheric dew point; the HTM says the test at the compressor plant confirms dryer performance, and failure to achieve the minimum dew point standard should not be accepted.
On Ultra Controlo equipment
The ULTRAAR ME has two integrated ULTRAGEST systems for continuous monitoring and control of the dew point. The ULTRAMED treatment unit accepts a digital dew point indicator for continuous outlet monitoring, which is not included as standard, and has a low dew point indicator alarm.
More questions
Is carbon monoxide monitored continuously at the air plant?
HTM 02-01 (2026) does not list carbon monoxide among the plant faults; it is covered by quality control testing of terminal units against a 5 ppm limit.
Who keeps the calibration records of contractor test equipment?
Under HTM 02-01 Part B, the contractor keeps the calibration records of its test equipment and produces them for inspection by the Authorised Person (MGPS).
Sources
- NHS England, HTM 02-01 Medical gas pipeline systems, Part A (2026), paras 7.68, 7.87, 7.89, 20.3, 20.38 to 20.40, 20.70, 20.72, 20.74, 20.75 and Table 33
- NHS England, HTM 02-01 Medical gas pipeline systems, Part B (2026), paras 11.67, 11.70, 11.71 and 11.178
- ULTRAAR ME datasheet, p. 2
- ULTRAMED datasheet, p. 3
Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.