PSA, cylinders or liquid oxygen: what is the difference?

Short answer

Under AIGA 113/20, oxygen 93 produced by PSA can be manufactured on site within the healthcare facility, while medicinal oxygen supplied in cylinders or as a cryogenic liquid is required to be manufactured by an approved supplier with a manufacturer’s licence.

Three ways to supply hospital oxygen

On-site PSA plantCylinders from a gas supplierLiquid oxygen (bulk tank)
Where it is madeOn site, from ambient air (WHO interim guidance, April 2020; International Pharmacopoeia, 2025)Manufactured by an approved supplier with a manufacturer’s licence, which AIGA 113/20 says is requiredAlways off site, by cryogenic distillation of ambient air (WHO interim guidance, April 2020; International Pharmacopoeia, 2025)
Who answers for its qualityQualified personnel of the healthcare facility, normally the head pharmacist (AIGA 113/20)The medical gas supplier (AIGA 113/20)The medical gas supplier (AIGA 113/20)
GradeOxygen 93: 90.0% to 96.0% oxygen, the rest mainly argon and nitrogen (International Pharmacopoeia, 2025)Above 99% oxygen (HTM 02-01, 2026)At least 99.5% oxygen (Ph. Eur. 0417; International Pharmacopoeia, Oxygen 99.5%)
Form at the hospitalGas, which can be piped directly to bedside terminal units (WHO interim guidance, April 2020)Gas at high pressure: the rated working pressure typically ranges from 137 to 150 bar, depending on the manufacturer (WHO, Foundations of medical oxygen systems, 2023)Liquid stored at down to −183 °C in a VIE (vacuum-insulated evaporator) and vaporised to gas for the pipeline (HTM 02-01, 2026)
Electricity at the hospitalNeeded: HTM 02-01 (2026) says all components of the PSA supply system should be connected to the essential electrical supplyNot needed: cylinders require no electricity to be functional (PATH and CHAI, 2021)Not needed: the tank feeds the pipeline by self-vaporisation (WHO interim guidance, April 2020)
Checks and monitoringDaily checklist, with oxygen purity recorded on the PLC and with a handheld analyser (Global Fund / BHI checklist, 2025)Periodic inspection and pressure testing at intervals set by national or international transport rules (see how often cylinders are retested)Telemetry should be installed so that the hospital and the gas supplier can monitor vessel levels and pressures continuously (HTM 02-01, 2026)

Sources

Ultra Controlo figures come from its technical datasheets. Third-party facts were read at the source on the date shown.

Reviewed: · Ultra Controlo engineering · Markdown version